Specifications

  • Category: Optical Emission Spectroscopy
  • Technique: ICP-OES/ ICP-AES
  • Manufacturer: Agilent
  • Country: USA
  • Available: Active
All specifications

About

Short description

ICP-OES with Vertical Dual View (VDV) configuration. Designed for routine and high-throughput labs. Features IntelliQuant software for advanced spectral interference detection and automation.

Specifications

ICP-OES with Vertical Dual View (VDV) configuration. Designed for routine and high-throughput labs. Features IntelliQuant software for advanced spectral interference detection and automation.

  • Status: Active
  • Country: USA
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Elemental coverage, including major and trace elements
  • Sensitivity : Sub-ppb detection limits
  • Dynamic range : 6–9 orders of magnitude dynamic range, depending on the configuration
  • Signal stability, relative standard deviation : Long-term stability with <1% RSD over 8 hours
  • Spray chamber : Glass cyclonic double-pass spray chamber; options for high-salt and HF applications
  • Nebulizer : Concentric glass nebulizer; optional configurations for ultrasonic or parallel-path nebulizers
  • Peristaltic pump : Fully PC-controlled peristaltic pump with variable speeds (0–80 rpm), up to five channels
  • RF generator frequency : 27 MHz
  • RF generator power : Adjustable from 700 to 1500 W in 10 W increments
  • View orientation/observation/optical system : Vertical Dual View (VDV) with radial and axial observations for flexibility
  • Detection system/detector : VistaChip II CCD detector
  • Grating /Wavelength range : 167–785 nm using echelle grating and single slit design for high resolution
  • Polychromator : Thermostatted polychromator with high thermal stability; uses CaF₂ prism cross-disperser
  • Typical resolution : <7 pm in the UV region
  • Gas supply: Argon ≥99.996% purity; mass flow-controlled plasma, auxiliary, nebulizer, and makeup gas flows
  • Power Requirements and Consumption : 200–240 VAC, 50/60 Hz; ~3 kVA
  • Environmental requirements : Operating temperature: 15–35°C; Humidity: 20–80% (non-condensing); Exhaust flow: 2.5 m³/min
  • Dimensions and Weight: 800 mm W × 740 mm D × 940 mm H; Weight: 106 kg

Articles

Other company product

Agilent 5973 inert

  • Category: Gas Chromatography - Mass Spectrometry
  • Technique: Single Quad GC-MS
  • Manufacturer: Agilent
  • Country: USA
  • Available: Discontinued
Detailed