Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Analytik Jena
  • Country: Germany
  • Available: Active
All specifications

About

Short description

Combines Flame and Graphite Furnace with automatic atomizer switching and advanced CS-AAS technology.

Specifications

Combines Flame and Graphite Furnace with automatic atomizer switching and advanced CS-AAS technology.

  • Status: Active
  • Country: Germany
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Across 72 elements, including molecular absorption lines
  • Signal stability, relative standard deviation : RSD <1% for repeat measurements
  • Atomizer types : Flame
  • Atomizer types : Furnace
  • Bandwidth : 0.002 nm at 200 nm; FWHM ≤ 3.8 pm for As 193.7 nm
  • Wavelength range : 185 to 900 nm
  • Wavelength accuracy : ±0.4 pm
  • Background correction : Iterative Baseline Correction (IBC) and Least Squares Background Correction (LSBC)
  • Optics type : High-resolution double monochromator with quartz prism and rotatable grating
  • Burner head material : Coded titanium burner head; adjustable rotation (0–90°)
  • Gas management : Comprehensive gas flow control for both Flame and Graphite Furnace
  • Detection system : CCD Array Detector
  • Monochromator : Double monochromator with high-resolution quartz optics
  • Lamp carousel : Not applicable; uses Xenon short arc lamp
  • Flame atomizer : Titanium burner; automatic ignition, extinguishing, and height adjustment
  • Furnace atomizer : Transversely Heated Graphite Furnace (THGA) with Stabilized Temperature Platform Furnace (STPF)
  • Furnace vision system : Integrated USB furnace camera for monitoring
  • Furnace heating rate/max temp : Up to 3000°C with a heating rate of 3000°C/s
  • Typical tube lifetime : Pyrolytically coated tubes; supports up to 50 µL sample volumes
  • Sensitivity (Furnace) : Pb detection: 0.66 µg/L at 283 nm
  • Sensitivity (Flame) : Cu detection: 0.015 mg/L at 324 nm (Air/Acetylene)
  • Safety features : Comprehensive safety interlocks for both Flame and Graphite Furnace systems
  • Power Requirements and Consumption : 230 V, 50/60 Hz; max 2100 VA
  • Environmental requirements : Operating temperature: 5–40°C; Humidity: 90% at 40°C non-condensing
  • Dimensions and Weight: 780 mm W × 775 mm D × 625 mm H; 170 kg

Articles

Other company product

novAA 800D

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Analytik Jena
  • Country: Germany
  • Available: Active
Detailed