Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Thermal Ionization Mass Spectrometry
  • Manufacturer: Kunyuan Instrument (Tianjin)
  • Country: China
  • Available: Active
All specifications

About

Short description

Thermal ionization mass spectrometer with double-focusing capabilities.

Specifications

Thermal ionization mass spectrometer with double-focusing capabilities.

  • Status: Active
  • Country: China
  • Isotopic: Yes
  • Type of measurable elements: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable isotopes/elements: 87Sr/86Sr, 143Nd/144Nd, 234U/238U, 235U/238U, and others
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Thermal ionization
  • Inlet types: filament drum with single- and dual-filament support
  • Mass range : 3-310 amu
  • Mass resolution/ Mass resolving power : >450 (5% valley definition)
  • Multicollector : Yes
  • Collectors qty : Up to 16 Faraday cups and 4-SEM channels (Secondary Electron Multipliers)
  • Abundance sensitivity : ≤0.1 ppm using RPQ (Retarding Potential Quadrupole) lens
  • Dynamic range : Wide dynamic range with simultaneous detection from cps to 50 V signal
  • Detection limits : Sub-ppm levels for isotope ratio measurements
  • Precision : High precision: ≤0.002% for 87Sr/86Sr and uranium isotopes (standard samples)
  • Signal stability, relative standard deviation : ±30 ppm/30 min
  • Detector type : Fixed multicollector array with Faraday cups and SEM; optimized with RPQ lens
  • Possible amplifiers : Up to 32 high-stability electrometric amplifiers with dynamic range up to 50 V; computer-controlled gain calibration
  • Acceleration voltage : 10 kV
  • Sample turret : 21-position filament drum with automated positioning
  • Geometry type : Double-focusing Nier-Johnson geometry
  • Focusing type : Multilens focusing with double quadrupole zoom optics
  • Magnetic sector radius/ angle : 250 mm; 90° deflection
  • Electric sector radius/ angle : 350 mm
  • Vacuum system : Oil-free system with 3 ion-getter pumps and a turbomolecular pump
  • Gas supply: Optional nitrogen bleed system for vacuum optimization
  • Power Requirements and Consumption : 100–240 V, 50/60 Hz; power consumption: 8 kVA
  • Environmental requirements : Temperature: 15 to 25°C; Humidity: ≤80%
  • Dimensions and Weight: Dimensions: 215 cm (L) x 184 cm (W) x 155 cm (H); Weight: 1500 kg