Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
All specifications

About

Short description

Single Flame Atomizer AAS with a fully automated gas box, high-resolution optics, and user-friendly SOLAAR software. Designed for routine laboratories requiring efficient flame analysis with occasional graphite furnace use.

Specifications

Single Flame Atomizer AAS with a fully automated gas box, high-resolution optics, and user-friendly SOLAAR software. Designed for routine laboratories requiring efficient flame analysis with occasional graphite furnace use.

  • Status: Active
  • Country: USA
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Over 70 elements, including heavy metals such as Pb, Cd, Zn, Cu, and Ni
  • Signal stability, relative standard deviation : RSD <0.5% for repeat analyses
  • Atomizer types : Flame
  • Bandwidth : Adjustable: 0.2, 0.5, 1.0 nm
  • Wavelength range : 180–900 nm
  • Wavelength accuracy : ±0.1 nm
  • Background correction : Guaranteed Quadline deuterium correction for high accuracy
  • Optics type : Double-beam optics for stability and low drift
  • Burner head material : Universal titanium burner for Air/Acetylene and N₂O/Acetylene flames
  • Gas management : Automatic binary control for precise flame conditions
  • Detection system : Photomultiplier tube (PMT)
  • Monochromator : 0.27 m Ebert type with high-resolution diffraction grating (1800 lines/mm)
  • Lamp carousel : 6-lamp coded, auto-aligning carousel
  • Photomultiplier : High-sensitivity PMT
  • Flame atomizer : Universal burner supporting both Air/Acetylene and N₂O/Acetylene flame types
  • Furnace atomizer : GFS33 Graphite Furnace module (optional) with dynamic optical feedback and rapid heating rates
  • Furnace vision system : Optional graphite furnace camera for precise injection monitoring
  • Sensitivity (Flame) : Cu detection: <0.003 µg/mL
  • Safety features : Integrated software and hardware safety interlocks, automatic flame shutdown, and gas leak detection
  • Power Requirements and Consumption : 230 V/110 V, 50/60 Hz, single phase; 500 W
  • Environmental requirements : Temperature: 10–35°C; Humidity: 20–80% (non-condensing)
  • Dimensions and Weight: 870 mm W × 620 mm D × 700 mm H; 115 kg

Articles

Other company product

Element XR

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: High resolution ICP-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Active
Detailed