Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
All specifications

About

Short description

Dedicated Graphite Furnace AAS with advanced Zeeman and D2 background correction, exceptional precision, and an integrated furnace vision system.

Specifications

Dedicated Graphite Furnace AAS with advanced Zeeman and D2 background correction, exceptional precision, and an integrated furnace vision system.

  • Status: Active
  • Country: USA
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Over 70 elements, including heavy metals such as Pb, Cd, Zn, and Cu
  • Signal stability, relative standard deviation : Relative Standard Deviation (RSD) <1% for trace element analysis
  • Atomizer types : Furnace
  • Bandwidth : 0.1, 0.2, 0.5, and 1.0 nm
  • Wavelength range : 180–900 nm
  • Wavelength accuracy : ±0.1 nm
  • Background correction : Guaranteed Quadline deuterium and Zeeman background correction for interference-free analysis
  • Optics type : Dual-beam optics with Echelle monochromator for high stability and low drift
  • Gas management : Automatic binary control for precise argon and protective gas flow regulation
  • Detection system : Photomultiplier tube (PMT)
  • Monochromator : High-resolution Echelle monochromator with sealed optics to minimize maintenance
  • Lamp carousel : Auto-aligning 6-lamp carousel with coded and uncoded lamp compatibility
  • Photomultiplier : Wide-range PMT for wavelengths from 180 to 900 nm
  • Furnace atomizer : GFS35(Z) module with dynamic optical temperature control and a max temperature of 3000°C
  • Furnace vision system : Integrated furnace camera for real-time sample monitoring during analysis
  • Furnace heating rate/max temp : Heating rate up to 3000°C/s; maximum temperature of 3000°C
  • Typical tube lifetime : Pyrolytically coated graphite tubes designed for extended lifetime, supporting up to 5000 firings
  • Sensitivity (Furnace) : Cd detection limit: <0.4 pg
  • Safety features : Integrated safety interlocks, automated gas shutoff, cuvette presence detection, and over-temperature protection
  • Power Requirements and Consumption : 230 V, 50/60 Hz, single phase; 460 VA
  • Environmental requirements : Temperature: 10–35°C; Humidity: 20–80% non-condensing
  • Dimensions and Weight: 870 mm W × 620 mm D × 700 mm H; weight 120 kg