Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Labtron
  • Country: China
  • Available: Active
All specifications

About

Short description

PC-controlled AAS with floated optics, 8-lamp stand, auto flame tuning.

Specifications

PC-controlled AAS with floated optics, 8-lamp stand, auto flame tuning.

  • Status: Active
  • Country: China
  • Country: UK
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Cu, Fe, etc.
  • Atomizer types : Flame
  • Atomizer types : Furnace
  • Bandwidth : 0.1–2.0 nm
  • Wavelength range : 190–900 nm
  • Wavelength accuracy : ≤ ±0.15 nm
  • Background correction : Yes, Deuterium + self-absorption
  • Optics type : Floated Optical Platform
  • Burner head material : Titanium
  • Gas management : Auto shutoff for pressure/flame
  • Detection system : Photomultiplier tube (PMT)
  • Monochromator : 1800 lines/mm
  • Lamp carousel : 8-lamp auto stand
  • Photomultiplier : -
  • Flame atomizer : Yes
  • Furnace atomizer : Yes
  • Furnace vision system : Yes
  • Furnace heating rate/max temp : Max 3000°C
  • Typical tube lifetime : -
  • Sensitivity (Furnace) : 0.004 µg/ml (Cu)
  • Sensitivity (Flame) : 0.02 µg/ml / 1% (Cu)
  • Safety features : Pressure drop, flameout, burner fault
  • Power Requirements and Consumption : 220 V, 50 Hz
  • Environmental requirements : -
  • Dimensions and Weight: 860 × 705 × 755 mm; 105 Kg

Articles

Other company product

LGC-B13

  • Category: Chromatography
  • Technique: Gas Chromatography
  • Manufacturer: Labtron
  • Country: China
  • Available: Active
Detailed