Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Labtron
  • Country: China
  • Available: Active
All specifications

About

Short description

Microcomputer-controlled AAS with vertical graphite furnace and automatic gas/temp control.

Specifications

Microcomputer-controlled AAS with vertical graphite furnace and automatic gas/temp control.

  • Status: Active
  • Country: China
  • Country: UK
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: -
  • Atomizer types : Flame
  • Atomizer types : Furnace
  • Bandwidth : 0.1, 0.2, 0.4, 0.7, 1.0, 1.4, 2.0 nm (auto)
  • Wavelength range : 185 – 900 nm
  • Wavelength accuracy : ± 0.15 nm
  • Background correction : -
  • Optics type : Full closed optical system
  • Burner head material : Full titanium, 50 or 100 mm
  • Gas management : Automatic gas control
  • Detection system : Photomultiplier tube (PMT)
  • Monochromator : -
  • Lamp carousel : ≤ 8 lamps, automatic alignment
  • Photomultiplier : Yes (imported)
  • Flame atomizer : Titanium combustion head
  • Furnace atomizer : Vertical heating graphite furnace
  • Furnace vision system : Vertical optical temperature monitoring
  • Furnace heating rate/max temp : RT to 3000°C
  • Typical tube lifetime : -
  • Sensitivity (Furnace) : 0.002 µg/mL (Cu)
  • Sensitivity (Flame) : 0.4 × 10⁻¹² g (Cd)
  • Safety features : Multiple alarms: temp, pressure, water flow
  • Power Requirements and Consumption : 110/220V ±5–10%, 60/50Hz; 5000 VA
  • Environmental requirements : -
  • Dimensions and Weight: -