Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Glow Discharge Mass Spectrometry
  • Manufacturer: Lumex
  • Country: Russia
  • Available: Active
All specifications

About

Short description

Time-of-Flight Mass Spectrometry (TOF-MS) with pulsed glow discharge plasma ionization. Direct analysis of solid-phase materials with elemental, isotopic and depth profiling.

Specifications

Time-of-Flight Mass Spectrometry (TOF-MS) with pulsed glow discharge plasma ionization. Direct analysis of solid-phase materials with elemental, isotopic and depth profiling.

  • Status: Active
  • Country: Russia
  • Isotopic: Yes
  • Elemental: Yes
  • Type of measurable elements: Most of elements, isotopes, and molecular species
  • Measurable isotopes/elements: Major, minor, and trace elements
  • Analyzer type : Time-of-Flight MS
  • Ion source type: Pulsed glow discharge plasma
  • Inlet types: Direct solid sample ablation or sputtering
  • Mass range : 1–2000 m/z
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : Up to 6000 (m/Δm at FWHM); nominal 4000
  • Sensitivity : 1–100 ppb
  • Detection limits : 1–100 ppb
  • Discharge parameters : Pulsed glow discharge; Ar < 1 L/h
  • Geometry type : TOF with reflectron
  • Focusing type : Gridless ion mirror (reflectron focusing)
  • Vacuum system : One forepump and two turbo molecular pumps
  • Gas supply: Argon < 1 L/h
  • Dimensions and Weight: 1450 × 800 × 1550 mm