Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Thermal Ionization Mass Spectrometry
  • Manufacturer: Finnigan MAT
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired high-precision thermal ionization mass spectrometer. Features a single-stage magnetic sector analyzer with stigmatic ion beam focusing

Specifications

Retired high-precision thermal ionization mass spectrometer. Features a single-stage magnetic sector analyzer with stigmatic ion beam focusing

  • Status: Discontinued
  • Country: USA
  • Isotopic: Yes
  • Type of measurable elements: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable isotopes/elements: Uranium isotopes, strontium, neodymium, lead, and other elements
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Thermal ionization
  • Inlet types: filament-based sample introduction system with sample turret
  • Mass range : 3-280 amu
  • Mass resolution/ Mass resolving power : Resolution at 10% peak height: 500
  • Multicollector : Yes
  • Collectors qty : Up to 9 Faraday collectors with one secondary electron multiplier (SEM)
  • Sensitivity : Optimized for trace uranium detection
  • Abundance sensitivity : <1 ppm for background ion signals
  • Precision : Dynamic mode: ±0.01% for 235U/238U; Static mode: ±0.002%
  • Signal stability, relative standard deviation : ±0.003% over 30 minutes
  • Detector type : Faraday cups and SEM for dynamic and static measurements
  • Possible amplifiers : High-stability amplifiers with noise suppression technology
  • Sample turret : 13-position automated sample turret
  • Geometry type : Single-stage double-focusing magnetic sector geometry
  • Focusing type : Stigmatic ion beam focusing
  • Magnetic sector radius/ angle : 640 mm; Deflection angle: 26.5°
  • Vacuum system : High-capacity oil-free turbo molecular pump and ion getter pumps
  • Power Requirements and Consumption : 220 V AC, ±15%, 50/60 Hz
  • Environmental requirements : Temperature: +18°C to +28°C; Humidity: 20–70% (non-condensing); Rate of temperature change: ≤1°C per hour
  • Dimensions and Weight: Dimensions: 1700 mm x 2400 mm x 1750 mm; Weight: 800 kg

Articles

Other company product

MAT 262

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Thermal Ionization Mass Spectrometry
  • Manufacturer: Finnigan MAT
  • Country: USA
  • Available: Discontinued
Detailed