Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Lumex
  • Country: -
  • Available: Active
All specifications

About

Short description

Zeeman Graphite Furnace AAS utilizing High Frequency Polarization Modulation (ZHFPM) for high selectivity and sensitivity. Compact design with quasi-double beam optics, eliminating drift. Designed for routine and advanced applications in environmental, food, and pharmaceutical analyses.

Specifications

Zeeman Graphite Furnace AAS utilizing High Frequency Polarization Modulation (ZHFPM) for high selectivity and sensitivity. Compact design with quasi-double beam optics, eliminating drift. Designed for routine and advanced applications in environmental, food, and pharmaceutical analyses.

  • Status: Active
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Over 70 elements, including heavy metals such as Pb, Cd, Zn, Cu, and Ni
  • Signal stability, relative standard deviation : -
  • Atomizer types : Furnace
  • Bandwidth : Not explicitly specified
  • Wavelength range : 185–900 nm
  • Wavelength accuracy : ±0.1 nm
  • Background correction : Zeeman High Frequency Polarization Modulation (ZHFPM) for effective background elimination
  • Optics type : Quasi-double beam optics with integrated drift correction
  • Gas management : Low argon consumption (<2 L/min) with automated flow control
  • Detection system : Photomultiplier tube (PMT)
  • Monochromator : Integrated high-resolution optics for narrow bandwidth measurements
  • Lamp carousel : Supports Hollow Cathode Lamps (HCL) and Electrodeless Discharge Lamps (EDL)
  • Photomultiplier : High-sensitivity PMT
  • Furnace atomizer : Stabilized Temperature Platform Furnace (STPF) for consistent and reliable atomization
  • Furnace vision system : -
  • Furnace heating rate/max temp : Heating rate up to 4500 K/s; maximum temperature of 3000°C
  • Typical tube lifetime : Up to 800 analysis cycles with standard graphite tubes
  • Sensitivity (Furnace) : Detection limits vary from <0.1 ppb to >1 ppb depending on the element
  • Safety features : Automated safety protocols for overheating, gas flow, and operational interlocks
  • Power Requirements and Consumption : Three-phase 220/380 VAC or 120/208 VAC, 50/60 Hz; 0.1 kW in standby, 6 kW in atomization mode
  • Environmental requirements : Operating temperature: 10–35°C; Humidity: 20–80% non-condensing
  • Dimensions and Weight: 800 mm L × 310 mm W × 475 mm H; weight 50 kg

Articles

Other company product

Lumas SOLID

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Glow Discharge Mass Spectrometry
  • Manufacturer: Lumex
  • Country: Russia
  • Available: Active
Detailed