Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Thermal Ionization Mass Spectrometry
  • Manufacturer: SELMI (ex Electron)
  • Country: Ukraine
  • Available: Discontinued
All specifications

About

Short description

Retired thermal ionization mass spectrometer. Capable of analyzing isotopes ranging from lithium to transuranic elements.

Specifications

Retired thermal ionization mass spectrometer. Capable of analyzing isotopes ranging from lithium to transuranic elements.

  • Status: Discontinued
  • Country: Ukraine
  • Isotopic: Yes
  • Type of measurable elements: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable isotopes/elements: Uranium isotopes: 234U, 235U, 236U, 238U and others
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Thermal ionization
  • Inlet types: Sample turret; dual-strip mode for simultaneous sample analysis
  • Mass range : 3 to 280 amu
  • Mass resolution/ Mass resolving power : >500 for uranium isotopes (5% valley definition)
  • Multicollector : Yes
  • Collectors qty : Five Faraday collectors with a secondary electron multiplier (SEM)
  • Abundance sensitivity : Isotopic sensitivity threshold for uranium <5 ppm (mass 237, Δm = 1)
  • Dynamic range : 5 x 10⁵ for Faraday detection; extended to 10⁸ in SEM mixed mode
  • Detection limits : Sub-ppm levels for isotopic measurements
  • Precision : Single-collector mode: <0.01% internal precision; Multi-collector mode: <0.002% internal precision
  • Signal stability, relative standard deviation : <0.005% over 20 minutes
  • Detector type : Faraday cups and secondary electron multiplier (SEM)
  • Possible amplifiers : High-stability amplifiers with 10¹¹ Ω resistors
  • Acceleration voltage : 10 kV
  • Sample turret : 22-position turret with motorized or manual adjustment
  • Geometry type : Double-focusing sector with a 90° magnetic analyzer
  • Focusing type : Stigmatic focusing
  • Magnetic sector radius/ angle : 200 mm radius; 90° deflection
  • Cooling system : Air-cooled
  • Vacuum system : Oil-free system: two charcoal sorption pumps, three ion-getter pumps, and a turbomolecular pump
  • Power Requirements and Consumption : Three-phase 220 V ±10%, 50 Hz; maximum power consumption: 5 kVA
  • Environmental requirements : Temperature: 10–35°C; Humidity: up to 80% at 25°C; Atmospheric pressure: 84.0–106.7 kPa
  • Dimensions and Weight: Analytical unit: 1500 x 1400 x 1500 mm; Operator desk: 1400 x 800 x 800 mm; Total weight: 2500 kg