Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Thermal Ionization Mass Spectrometry
  • Manufacturer: Uralpribor
  • Country: Russia
  • Available: Active
All specifications

About

Short description

High-precision thermal ionization mass spectrometer (TIMS) designed for isotopic analysis of uranium in solid-phase samples. Provides automated sample preparation, high resolution, and robust performance for nuclear and geochemical applications.

Specifications

High-precision thermal ionization mass spectrometer (TIMS) designed for isotopic analysis of uranium in solid-phase samples. Provides automated sample preparation, high resolution, and robust performance for nuclear and geochemical applications.

  • Status: Active
  • Country: Russia
  • Isotopic: Yes
  • Type of measurable elements: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable isotopes/elements: 234U, 235U, 236U, 238U
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Thermal ionization
  • Inlet types: Filament loading system with manual or automated adjustment
  • Mass range : Up to 300 amu
  • Mass resolution/ Mass resolving power : 750
  • Multicollector : Yes
  • Collectors qty : 10
  • Sensitivity : optimized for trace-level detection uranium isotopes
  • Abundance sensitivity : ≤ 5 ppm for 238U
  • Precision : ±0.04% for 235U in the range 0.5%−1.0% for 234U/236U at 0.0005%−0.006%
  • Signal stability, relative standard deviation : ≤ 20 ppm over 20 minutes for 238U
  • Detector type : Faraday cups and secondary electron multiplier (SEM)
  • Possible amplifiers : High-stabilitHighly sensitive input amplifier in a special vacuum chamber with suppressed microphone effect, ensuring measurement accuracy.y amplifiers with reduced noise for ultra-precise measurements
  • Geometry type : Double-focusing geometry with magnetic sector
  • Focusing type : Electrostatic focusing system
  • Power Requirements and Consumption : 220 V ±10%, 50 Hz; Maximum power consumption: 3 kVA
  • Environmental requirements : Temperature: 10–35°C; Humidity: ≤ 80%; Atmospheric pressure: 84–106.7 kPa
  • Dimensions and Weight: Analytical unit: 1800×1450×1800 mm; Electronics unit: 1200×650×1700 mm; Operator desk: 1500×1250×850 mm; Weight: 1400 kg

Articles

Other company product

MTI 350-G

  • Category: Other Types of IOMS
  • Technique: Multicollector Mass Spectrometry for Uranium Hexafluoride
  • Manufacturer: Uralpribor
  • Country: Russia
  • Available: Discontinued
Detailed