Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Secondary Ion Mass Spectrometry
  • Manufacturer: Compact Science Systems
  • Country: France
  • Available: Active
All specifications

About

Short description

Next-gen NanoSIMS with enhanced lateral resolution, higher throughput, and advanced sample handling features. Ideal for high-precision nanoanalysis.

Specifications

Next-gen NanoSIMS with enhanced lateral resolution, higher throughput, and advanced sample handling features. Ideal for high-precision nanoanalysis.

  • Status: Active
  • Country: France
  • Isotopic: Yes
  • Elemental: Yes
  • Type of measurable elements: Most of elements, isotopes, and molecular species
  • Measurable isotopes/elements: Isotopes including H, C, N, O, Si, and others with superior isotopic ratio capabilities
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Advanced Cs+ and redesigned RF-plasma ion source
  • Inlet types: Larger sample holders for diverse applications
  • Mass range : given by the factor D = X 22 between minimum mass and maximum mass (ex: from 1 amu to 22 amu or from 10 amu to 220 amu depending on the selected B-field in the electro- magnet
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : Up to 7500 (depending on slit size and configuration)
  • Multicollector : Yes
  • Collectors qty : 7 with enhanced collection capabilities
  • Detection limits : Sub-ppb detection for select isotopes in ideal conditions
  • Detector type : Electron multipliers and Faraday cups
  • Primary beam : Improved Cs+ and O- with optimized source stability
  • Voltage : +8 kV (Cs+), -8 kV (O-)
  • Spatial resolution : ≤ 30 nm (Cs+), ≤ 50 nm (O-)
  • Depth profile : Improved to ≤ 12 nm per decade with low-energy Cs+ beams
  • Beam energy : Adjustable, with improved depth profiling precision
  • Camera : Enhanced optical imaging for better accuracy
  • Geometry type : Coaxial, optimized for secondary ion collection
  • Cooling system : Water cooling for sources and detectors
  • Vacuum system : Turbo-molecular and ion pumps
  • Gas supply: Standard gas delivery systems for Cs and O2
  • Dimensions and Weight: L: 6.8 m, W: 5.2 m, H: 2.85 m