Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Analytik Jena
  • Country: Germany
  • Available: Active
All specifications

About

Short description

Combines Flame and Graphite Furnace modes with automated atomizer switching and advanced safety interlocks.

Specifications

Combines Flame and Graphite Furnace modes with automated atomizer switching and advanced safety interlocks.

  • Status: Active
  • Country: Germany
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Over 70 elements, including Pb, Cd, Zn, and Cu
  • Signal stability, relative standard deviation : RSD <1% for routine analyses
  • Atomizer types : Flame
  • Atomizer types : Furnace
  • Bandwidth : Adjustable slit widths: 0.2, 0.3, 0.5, 0.8, 1.2 nm
  • Wavelength range : 185–900 nm
  • Wavelength accuracy : ±0.1 nm
  • Background correction : Deuterium background correction with D2 hollow cathode lamp (HCL)
  • Optics type : Czerny-Turner optical design with purgeable optics for stability
  • Burner head material : Coded titanium burner head with automated rotation and height adjustment
  • Gas management : Comprehensive gas control for both flame and graphite furnace modes
  • Detection system : SiOSens solid-state detector
  • Monochromator : Holographic grating with 1800 lines/mm; rotatable for precise wavelength adjustment
  • Lamp carousel : 8-fold turret with automatic RFID recognition
  • Flame atomizer : Supports Air/Acetylene and N₂O/Acetylene configurations
  • Furnace atomizer : Stabilized Temperature Platform Furnace (STPF) with max temperature of 3000°C
  • Furnace vision system : Integrated USB camera for real-time monitoring of injection and atomization processes
  • Furnace heating rate/max temp : Heating rate up to 3000°C/s; maximum temperature of 3000°C
  • Typical tube lifetime : Pyrolytically coated graphite tubes designed for extended lifetime
  • Sensitivity (Furnace) : Pb detection: 0.8 µg/L (20 µL sample)
  • Sensitivity (Flame) : Cu detection: 0.02 mg/L with 100 mm burner head
  • Safety features : Comprehensive safety interlocks for both atomization modes
  • Power Requirements and Consumption : 230 V, 50/60 Hz, max 1400 VA
  • Environmental requirements : Operating temperature: 5–40°C; Humidity: 90% at 40°C non-condensing
  • Dimensions and Weight: 820 mm W × 770 mm D × 600 mm H; 130 kg