Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: Analytik Jena
  • Country: Germany
  • Available: Active
All specifications

About

Short description

Flame AAS optimized for routine analyses with coded burner heads and automated features for high reliability.

Specifications

Flame AAS optimized for routine analyses with coded burner heads and automated features for high reliability.

  • Status: Active
  • Country: Germany
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Over 70 elements, including Pb, Cd, Zn, and Cu
  • Signal stability, relative standard deviation : RSD <1% for routine analyses
  • Atomizer types : Flame
  • Bandwidth : Adjustable slit widths: 0.2, 0.3, 0.5, 0.8, 1.2 nm
  • Wavelength range : 185–900 nm
  • Wavelength accuracy : ±0.1 nm
  • Background correction : Deuterium background correction with D2 hollow cathode lamp (HCL)
  • Optics type : Czerny-Turner optical design with purgeable optics for stability
  • Burner head material : Coded titanium burner head with automated rotation and height adjustment
  • Gas management : Automated gas management system with multiple safety interlocks
  • Detection system : SiOSens solid-state detector
  • Monochromator : Holographic grating with 1800 lines/mm; rotatable for precise wavelength adjustment
  • Lamp carousel : 8-fold turret with automatic RFID recognition
  • Flame atomizer : Supports Air/Acetylene and N₂O/Acetylene configurations
  • Sensitivity (Flame) : Cu detection: 0.02 mg/L with 100 mm burner head
  • Safety features : Multiple sensors monitoring burner head, siphon system, and gas management system
  • Power Requirements and Consumption : 230 V/110 V, 50/60 Hz, 150 VA
  • Environmental requirements : Operating temperature: 5–40°C; Humidity: 90% at 40°C non-condensing
  • Dimensions and Weight: 820 mm W × 770 mm D × 600 mm H; 95 kg