Specifications

  • Category: Optical Emission Spectroscopy
  • Technique: Spark-OES/GD-OES
  • Manufacturer: Skyray
  • Country: China
  • Available: -
All specifications

About

Short description

Multi-channel spark-OES designed for rapid quantitative analysis of ferrous and non-ferrous metals, suitable for quality control in metallurgy, machinery, and other industries.

Specifications

Multi-channel spark-OES designed for rapid quantitative analysis of ferrous and non-ferrous metals, suitable for quality control in metallurgy, machinery, and other industries.

  • Country: China
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Capable of analyzing elements such as Al, Pb, Mg, Zn, Sn, Fe, Co, Ni, Ti, Cu, and others.
  • Sensitivity : High sensitivity with excellent repeatability and stability; specific detection limits -.
  • Possible Matrixes : Suitable for various metals and alloys, including ferrous and non-ferrous metals.
  • Optic System : Paschen-Runge optical system with a concave grating (radius of curvature: 998.8 mm; etching density: 2160 L/mm).
  • Optical chamber : Vacuum optical chamber with oil return prevention technology; maintained at a constant temperature of 38°C ± 0.1°C.
  • Detection system/detector : Multi-channel detection system supporting up to 50 channels; specific detector type -.
  • Grating /Wavelength range : Wavelength range: 170–463 nm; reciprocal linear dispersion: 0.46 nm/mm; entry slit width: 20 μm; exit slit width: 70 μm.
  • Spark Stand : Argon-flushed stimulation chamber designed to minimize argon gas consumption.
  • Source frequency : Adjustable spark frequency ranging from 150 to 600 Hz, depending on the material.
  • Spark Stand : Stable spark power supply with a stimulation power of 15 kV.
  • Correction : Features include data processing control using single-chip computer and FPGA; specific correction methods not detailed.
  • Vacuum system : Yes; vacuum maintained at less than 10 Pa within the optical chamber.
  • Gas supply: Requires argon gas for the sparking process; designed for minimal consumption.

Articles

Other company product

OES100VMI 

  • Category: Optical Emission Spectroscopy
  • Technique: Spark-OES/GD-OES
  • Manufacturer: Skyray
  • Country: China
  • Available: -
Detailed