Specifications

  • Category: Optical Emission Spectroscopy
  • Technique: Spark-OES/GD-OES
  • Manufacturer: PG Instruments
  • Country: UK
  • Available: Active
All specifications

About

Short description

CCD-based OES for metals with full-spectrum scanning and excellent LOD for trace elements

Specifications

CCD-based OES for metals with full-spectrum scanning and excellent LOD for trace elements

  • Status: Active
  • Country: UK
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: C, N, S, P, B, As, Li, Na, heavy and trace metals across Fe, Al, Cu, Ti, Mg, etc.
  • Sensitivity : 10 ppm or lower (C, N, S, As, etc.); ultra-trace detection possible
  • Possible Matrixes : Fe, Al, Cu, Ni, Co, Mg, Ti, Zn, Pb, Ag, Sn, and more
  • Optic System : Paschen-Runge spectrometer; 500 mm focal length; 2700 lines/mm holographic grating
  • Optical chamber : Argon-flushed chamber with thermal insulation and software-controlled gas system
  • Detection system/detector : High-resolution linear CCD (3648 pixels, 8 μm)
  • Grating /Wavelength range : 130–900 nm
  • Spark Stand : Self-purging coaxial spark stand; max sample weight 50 kg
  • Source frequency : Up to 1000 Hz
  • Correction : Matrix correction, single-sample recalibration, automatic element form correction
  • Gas supply: Argon, <8 L/min during analysis; 60 mL/min standby; software-controlled
  • Power Requirements and Consumption : 220 V ±10%, 16 A, 2.5 kVA, single phase
  • Environmental requirements : 10–50 °C, <75% RH
  • Dimensions and Weight: 840 × 470 × 440 mm; ~80 kg

Articles

Other company product

AA500F

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: PG Instruments
  • Country: UK
  • Available: Active
Detailed