Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: PerkinElmer
  • Country: USA
  • Available: Active
All specifications

About

Short description

Flame-only AAS designed for routine analyses, offering true double-beam optics, high sensitivity, and Syngistix software for simple workflows.

Specifications

Flame-only AAS designed for routine analyses, offering true double-beam optics, high sensitivity, and Syngistix software for simple workflows.

  • Status: Active
  • Country: USA
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Over 70 elements, including Pb, Cd, Zn, Cu, and Ni
  • Signal stability, relative standard deviation : RSD <0.3% for flame measurements
  • Atomizer types : Flame
  • Bandwidth : User-selectable slit widths: 0.2, 0.7, and 2.0 nm
  • Wavelength range : 184–900 nm
  • Wavelength accuracy : ±0.1 nm
  • Background correction : Deuterium arc lamp for accurate flame background correction
  • Optics type : True double-beam optics with fiber optics for enhanced detection limits
  • Burner head material : Titanium burner head, corrosion-resistant; standard 10-cm single-slot configuration
  • Gas management : Fully computer-controlled with oxidant and fuel monitoring
  • Detection system : SSD (segmented, CMOS)
  • Monochromator : Littrow design with motorized drive for automatic wavelength selection and peaking
  • Lamp carousel : Motorized 8-lamp mount for Lumina™ hollow cathode lamps and electrodeless discharge lamps (EDLs)
  • Flame atomizer : Air/Acetylene and N₂O/Acetylene configurations with modular sample introduction system
  • Sensitivity (Flame) : Cu detection limit: <0.004 µg/mL
  • Safety features : Interlocks for burner position, gas pressure, liquid drain system, and flame monitoring
  • Power Requirements and Consumption : 230 V, 50/60 Hz; max 800 VA
  • Environmental requirements : Ambient temperature: +10 °C to +35 °C Relative humidity: 20 to 80% non-condensing
  • Dimensions and Weight: 95 cm (37.5 in.) 73 cm (29 in.) 68 cm (27 in.) 94 kg (207 lb)