Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: PerkinElmer
  • Country: USA
  • Available: Active
All specifications

About

Short description

Combines Flame and HGA capabilities with automated switching and enhanced safety features for precise multi-mode analyses.

Specifications

Combines Flame and HGA capabilities with automated switching and enhanced safety features for precise multi-mode analyses.

  • Status: Active
  • Country: USA
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Over 70 elements, including Pb, Cd, Zn, Cu, and Ni
  • Signal stability, relative standard deviation : RSD <0.3% for flame measurements; <0.5% for furnace measurements
  • Atomizer types : Flame
  • Atomizer types : Hydride Generation
  • Bandwidth : User-selectable slit widths: 0.2, 0.7, and 2.0 nm
  • Wavelength range : 184–900 nm
  • Wavelength accuracy : ±0.1 nm
  • Background correction : Deuterium arc for flame; longitudinal Zeeman-effect for furnace analyses
  • Optics type : True double-beam optics with fiber optics for enhanced detection limits
  • Burner head material : Titanium burner head, corrosion-resistant; standard 10-cm single-slot configuration
  • Gas management : Integrated gas management for both modes
  • Detection system : SSD (segmented, CMOS)
  • Monochromator : Littrow design with motorized drive for automatic wavelength selection and peaking
  • Lamp carousel : Motorized 8-lamp mount for Lumina™ hollow cathode lamps and electrodeless discharge lamps (EDLs)
  • Flame atomizer : Air/Acetylene and N₂O/Acetylene configurations with modular sample introduction system
  • Furnace atomizer : Heated Graphite Furnace Atomizer (HGA)
  • Furnace vision system : Built-in TubeView color furnace camera for sample monitoring
  • Furnace heating rate/max temp : Heating rate up to 2600°C with controlled temperature profiles
  • Typical tube lifetime : Extended lifetime due to optimized heating profiles
  • Sensitivity (Furnace) : Cd detection limit: <0.2 pg
  • Sensitivity (Flame) : Cu detection limit: <0.004 µg/mL
  • Safety features : Comprehensive safety interlocks for both flame and furnace operations
  • Power Requirements and Consumption : 230 V, 50/60 Hz; max 7000 VA
  • Environmental requirements : Ambient temperature: +10 °C to +35 °C Relative humidity: 20 to 80% non-condensing
  • Dimensions and Weight: 95 cm (37.5 in.) 73 cm (29 in.) 68 cm (27 in.) 123 kg (271 lb)