Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Thermal Ionization Mass Spectrometry
  • Manufacturer: Micromass
  • Country: UK
  • Available: Discontinued
All specifications

About

Short description

Retired thermal ionization mass spectrometer.

Specifications

Retired thermal ionization mass spectrometer.

  • Status: Discontinued
  • Country: UK
  • Isotopic: Yes
  • Type of measurable elements: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable isotopes/elements: 234U, 235U, 236U, 238U, 87Sr/86Sr, 143Nd/144Nd, and others
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Thermal ionization
  • Inlet types: sample turret; optional filament preheating and oxygen bleed system
  • Mass range : 3-280 amu
  • Mass resolution/ Mass resolving power : > 450
  • Multicollector : Yes
  • Collectors qty : Up to 9 motorized Faraday collectors; optional Daly detector or electron multiplier
  • Dynamic range : Wide dynamic range for isotope measurements
  • Detection limits : Sub-ppm levels for trace isotope analysis
  • Precision : High-precision isotope ratio measurements
  • Detector type : Long-lifetime Faraday cups and ion counting Daly detector
  • Possible amplifiers : Cooled amplifiers for high precision
  • Acceleration voltage : 10 kV
  • Sample turret : 20-position sample turret
  • Geometry type : Double-focusing magnetic sector geometry
  • Focusing type : ion optics with adjustable collector slits
  • Magnetic sector radius/ angle : 200 mm radius, 90° deflection
  • Cooling system : Cooled amplifiers and source components
  • Vacuum system : High-capacity vacuum system with bake-out ovens
  • Gas supply: Oxygen bleed system for optimized ionization
  • Power Requirements and Consumption : 220 V ±20%, 50/60 Hz; Power consumption: 7.5 kW
  • Environmental requirements : Temperature: 18–24°C; Humidity: 20–80%; Atmospheric pressure: 84–106.7 kPa
  • Dimensions and Weight: 1800 x 1000 x 1750 mm; Weight: 1000 kg