Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Secondary Ion Mass Spectrometry
  • Manufacturer: Australian Scientific Instruments
  • Country: Australia
  • Available: Discontinued
All specifications

About

Short description

Retired high-precision SIMS for isotopic and elemental analysis. Optimized for U-Pb dating and trace elements.

Specifications

Retired high-precision SIMS for isotopic and elemental analysis. Optimized for U-Pb dating and trace elements.

  • Status: Discontinued
  • Country: Australia
  • Isotopic: Yes
  • Elemental: Yes
  • Type of measurable elements: Most of elements, isotopes, and molecular species
  • Measurable isotopes/elements: Isotopes of U, Pb, Th, and trace elements
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Cs+ and O2- primary ion beams
  • Inlet types: Solid sample mounts
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : Up to 13,000
  • Sensitivity : Better than 18 cps/ppm for Pb isotopes
  • Detection limits : Sub-ppm level
  • Detector type : Electron multiplier and Faraday cup
  • Primary beam : Cs+ or O2-
  • Voltage : ~10–15 kV
  • Spatial resolution : 5–30 µm
  • Depth profile : 1–5 µm sputtering depth
  • Beam energy : ~10–15 kV
  • Camera : Integrated CCD camera
  • Geometry type : Forward geometry
  • Focusing type : Magnetic and electrostatic focusing
  • Magnetic sector radius/ angle : 1000 mm/ 72.5 deg
  • Electric sector radius/ angle : 1270 mm/ 85 deg
  • Cooling system : Active and passive cooling
  • Vacuum system : Differential pumping
  • Dimensions and Weight: Beamline >7 m, weight >12 tonnes