Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Secondary Ion Mass Spectrometry
  • Manufacturer: Australian Scientific Instruments
  • Country: Australia
  • Available: Discontinued
All specifications

About

Short description

Retired specialized SIMS optimized for stable isotope analysis of light elements such as O, C, and S.

Specifications

Retired specialized SIMS optimized for stable isotope analysis of light elements such as O, C, and S.

  • Status: Discontinued
  • Country: Australia
  • Isotopic: Yes
  • Elemental: Yes
  • Type of measurable elements: Light stable isotopes (C, N, S, O, H)
  • Measurable isotopes/elements: Light isotopes of O, C, S, and others
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Cs+ and O2- primary ion beams
  • Inlet types: Solid sample mounts
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : Up to 13,000
  • Multicollector : Yes
  • Collectors qty : 3 (low mass, axial, high mass)
  • Detection limits : Sub-ppm for light isotopes
  • Precision : <0.3% for stable isotope ratios
  • Detector type : Electron multipliers with low-noise amplification
  • Possible amplifiers : Enhanced amplifiers with improved thermal stability
  • Primary beam : Cs+ or O2-
  • Voltage : ~10–15 kV
  • Spatial resolution : 5–30 µm
  • Depth profile : 1–5 µm sputtering depth
  • Beam energy : ~10–15 kV
  • Camera : Integrated CCD camera
  • Geometry type : Forward geometry
  • Focusing type : Magnetic and electrostatic focusing
  • Magnetic sector radius/ angle : 1000 mm/ 72.5 deg
  • Electric sector radius/ angle : 1270 mm/ 85 deg
  • Cooling system : Active and passive cooling
  • Vacuum system : Differential pumping with enhanced stable isotope configurations

Articles

Other company product

SHRIMP IIe

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Secondary Ion Mass Spectrometry
  • Manufacturer: Australian Scientific Instruments
  • Country: Australia
  • Available: Discontinued
Detailed