Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Thermal Ionization Mass Spectrometry
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Active
All specifications

About

Short description

Thermal ionization mass spectrometer designed for precision and sensitivity. Incorporates 1013 Ω Amplifier Technology, multicollector flexibility, and RPQ lenses for high abundance sensitivity.

Specifications

Thermal ionization mass spectrometer designed for precision and sensitivity. Incorporates 1013 Ω Amplifier Technology, multicollector flexibility, and RPQ lenses for high abundance sensitivity.

  • Status: Active
  • Country: Germany
  • Isotopic: Yes
  • Type of measurable elements: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable isotopes/elements: 87Sr/86Sr, 143Nd/144Nd, 234U/238U, 206Pb/204Pb, and others
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Thermal ionization
  • Inlet types: filament sample turret with single and double filament options, optional preheat system
  • Mass range : 3-310 amu
  • Mass resolution/ Mass resolving power : >450 (10% valley definition)
  • Multicollector : Yes
  • Collectors qty : 9 Faraday cups (movable) and up to 8 ion counters (SEM/CDDs)
  • Sensitivity : High sensitivity with dynamic range from 1 cps to 30 Gcps using 10¹³ Ω amplifiers and SEM detectors
  • Abundance sensitivity : Standard: <2 ppm at 238U; With RPQ lenses: <2 × 10⁻⁸ for extreme isotope ratios
  • Dynamic range : 12 orders of magnitude; 1 cps−30 Gcps
  • Detection limits : Sub-ppm levels for trace isotopes
  • Precision : Better than 0.01% for U-Pb geochronology; 87Sr/86Sr: ±0.005‰
  • Signal stability, relative standard deviation : <0.005% over extended measurement periods
  • Detector type : Faraday cups, discrete dynode Secondary Electron Multipliers (SEM), and Compact Discrete Dynode Detectors (CDD)
  • Possible amplifiers : 10¹¹ Ω (standard), 10¹³ Ω Amplifier Technology; software-controlled gain calibration with tau correction
  • Acceleration voltage : 10 kV
  • Sample turret : 21-position turret with motorized and manual adjustment options
  • Geometry type : Double-focusing magnetic sector geometry
  • Focusing type : Stigmatic focusing with dynamic zoom optics
  • Magnetic sector radius/ angle : 230 mm radius; 90° deflection
  • Cooling system : Water-cooled magnet system; air-cooled electronics
  • Vacuum system : Turbo molecular pumps, ion-getter pumps, cryogenic pump options for rapid pump down
  • Gas supply: Optional gas bleed valve for specific isotopic systems
  • Power Requirements and Consumption : 230 V, 50/60 Hz
  • Environmental requirements : Operating temperature: 18–25°C; Humidity: <60%

Articles

Other company product

Phenom Perception GSR

  • Category: Electron Microscopy
  • Technique: Desctop EM
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
Detailed