Specifications

  • Category: Spectroscopy
  • Technique: Atomic Absorption Spectroscopy
  • Manufacturer: GBC Scientific Equipment
  • Country: Australia
  • Available: Active
All specifications

About

Short description

XplorAA Atomic Absorption Spectrometer; Economical workhorse for versatile analysis; Available in flame, no-flame, or hybrid configurations.

Specifications

XplorAA Atomic Absorption Spectrometer; Economical workhorse for versatile analysis; Available in flame, no-flame, or hybrid configurations.

  • Status: Active
  • Country: Australia
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Mercury analysis down to parts per trillion levels; Hydride-forming elements like arsenic, selenium, antimony, bismuth, etc.
  • Signal stability, relative standard deviation : ≤ 0.5% RSD for flame; - for furnace
  • Atomizer types : Flame
  • Atomizer types : Furnace
  • Atomizer types : Hydride Generation
  • Bandwidth : Adjustable continuously: 0.1 to 2.0 nm (0.1 nm increments)
  • Wavelength range : 175–900 nm
  • Wavelength accuracy : ±0.10 nm
  • Background correction : Hyper-Pulse background corrector, Deuterium arc lamp for 175–425 nm; Corrects up to 2.5 absorbance
  • Optics type : Double or single beam optics; Asymmetric modulation; Quartz-coated reflective mirrors
  • Burner head material : All-titanium for corrosion resistance
  • Gas management : Programmable, automatic, or manual gas control options with safety interlocks
  • Detection system : Photomultiplier tube (PMT)
  • Monochromator : Ebert-Fastie design; 333 mm focal length; Holographic grating with 1800 lines/mm
  • Lamp carousel : Two-lamp turret (manual) or six-lamp turret (automatic); Lamp optimization and alignment automation
  • Photomultiplier : Multi-alkali wide-range photomultiplier tube
  • Flame atomizer : Premix system with safety interlocks; Adjustable sample uptake rate; Air-acetylene and nitrous oxide-acetylene flame capabilities
  • Furnace atomizer : Graphite furnace with heating up to 3000°C; Fully automated with PAL auto sampler support
  • Furnace vision system : -
  • Furnace heating rate/max temp : Up to 3000°C; Heating rate up to 2000°C/sec
  • Typical tube lifetime : -
  • Sensitivity (Furnace) : Sub-ppb detection limits depending on element and conditions
  • Sensitivity (Flame) : > 0.9 absorbance for 5 mg/L copper solution with RSD < 0.5%
  • Safety features : Flame safety interlocks for gas pressure, burner presence, liquid trap levels, oxidant flow, and ignition conditions
  • Power Requirements and Consumption : 100–240 V AC, 50/60 Hz
  • Environmental requirements : Protected against harsh environments with optional air purge for electronics
  • Dimensions and Weight: 820 x 520 x 480 mm (W x D x H); Weight: 67 kg unpacked, 100 kg packed

Articles

Other company product

Quantima

  • Category: Optical Emission Spectroscopy
  • Technique: ICP-OES/ ICP-AES
  • Manufacturer: GBC Scientific Equipment
  • Country: Australia
  • Available: Active
Detailed