Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Thermal Ionization Mass Spectrometry
  • Manufacturer: MSTech Group
  • Country: China
  • Available: Active
All specifications

About

Short description

High-precision double-focusing thermal ionization mass spectrometer (TIMS) with fixed multicollectors and variable dispersion optics.

Specifications

High-precision double-focusing thermal ionization mass spectrometer (TIMS) with fixed multicollectors and variable dispersion optics.

  • Status: Active
  • Country: China
  • Isotopic: Yes
  • Type of measurable elements: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable isotopes/elements: Lithium to uranium isotopes, including 6Li/7Li, 234U/238U, and other actinides
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Thermal ionization
  • Inlet types: filament sample turret with single and double filament options, optional preheat system
  • Mass range : 3-280 amu
  • High Resolution : Yes
  • Multicollector : Yes
  • Collectors qty : Up to 15 fixed-position Faraday collectors and up to 4 full-size discrete dynode electron multipliers (SEM)
  • Sensitivity : High sensitivity; 238U isotopic sensitivity: ≤2 ppm (without RPQ), ≤0.1 ppm (with RPQ lens)
  • Abundance sensitivity : ≤0.1 ppm with RPQ lens for uranium isotopes
  • Dynamic range : dynamic range from 0 to 50V
  • Detection limits : Sub-ppm detection limits
  • Precision : Internal precision: <0.002% for 238U (U500 standard)
  • Signal stability, relative standard deviation : ±30 ppm over 30 minutes
  • Detector type : Faraday cups and Secondary Electron Multipliers (SEM) with RPQ lens for improved sensitivity
  • Possible amplifiers : 15 amplifiers, noise ≤ 40 μVrms (10¹¹ Ω, 4-second integration time)
  • Sample turret : 21-position automated cathode drum
  • Geometry type : Nier-Johnson straight geometry with double focusing
  • Focusing type : Variable dispersion zoom optics using dual quadrupole lenses
  • Magnetic sector radius/ angle : Magnetic analyzer radius: 350 mm; 90° deflection
  • Electric sector radius/ angle : Electrostatic analyzer radius: 250 mm
  • Vacuum system : Static vacuum system, UHV ≤ 5 × 10^-8 mbar in the ion source, ≤ 5 × 10^-9 mbar in the analyzer Oil-free vacuum system with forevacuum, turbomolecular pump, and nitrogen trap
  • Gas supply: Optional nitrogen bleed system