Specifications

  • Category: Isotopic Spectroscopy
  • Technique: Tunable infrared laser direct absorption spectroscopy
  • Manufacturer: Aerodyne Research
  • Country: USA
  • Available: Active
All specifications

About

Short description

Dual-laser analyzer for direct spectroscopic measurement of methane isotopes without chemical separation.

Specifications

Dual-laser analyzer for direct spectroscopic measurement of methane isotopes without chemical separation.

  • Status: Active
  • Country: USA
  • Isotopic: Yes
  • Type of measurable elements: Light stable isotopes (C, N, S, O, H)
  • Measurable isotopes/elements: CH₄, δ¹³CH₄, δCH₃D
  • Dynamic range : 2–20 ppm
  • Precision : Continuous flow (air): 1 s — CH₄ 0.2 ppb, δ¹³CH₄ 1 ‰, δCH₃D 30 ‰; 100 s — CH₄ 0.05 ppb, δ¹³CH₄ 0.25 ‰, δCH₃D 8 ‰. Discrete samples: 125 mL (3 min) — δ¹³CH₄ 0.3 ‰, δCH₃D 10 ‰; 10×125 mL (30 min) — δ¹³CH₄ 0.1 ‰, δCH₃D 3 ‰.
  • Cooling system : Thermoelectric chiller
  • Vacuum system : Vacuum pump (customer specified)
  • Gas supply: Flushing optics with CO₂-free gas recommended
  • Power Requirements and Consumption : 250–500 W; 120/240 V; 50/60 Hz (without pump)
  • Environmental requirements : Operating −20 to 40 °C
  • Dimensions and Weight: 56 × 77 × 64 cm; 75 kg