Specifications

  • Category: Cavity-Enhanced Spectroscopy (isotopic)
  • Technique: Cavity ring-down spectroscopy
  • Manufacturer: Picarro
  • Country: USA
  • Available: Active
All specifications

About

Short description

Analyzer for δ¹³C isotopes in CO₂ and CH₄. Operates in three modes: CO₂-only, CH₄-only, and combined CO₂-CH₄ mode. Ideal for field and laboratory applications with low maintenance requirements.

Specifications

Analyzer for δ¹³C isotopes in CO₂ and CH₄. Operates in three modes: CO₂-only, CH₄-only, and combined CO₂-CH₄ mode. Ideal for field and laboratory applications with low maintenance requirements.

  • Status: Active
  • Country: USA
  • Isotopic: Yes
  • Type of measurable elements: Light stable isotopes (C, N, S, O, H)
  • Measurable isotopes/elements: δ¹³C in CO₂ and CH₄
  • Precision : δ¹³C-CO₂: <0.12‰ (CO₂-only); δ¹³C-CH₄: <0.8‰ (High Precision mode), <0.4‰ (High Dynamic Range mode) CO₂: ±200 ppb; CH₄: ±50 ppb
  • Drift : δ¹³C-CO₂: <0.6‰; δ¹³C-CH₄: <1.15‰ over 24 hrs (peak-to-peak, 1 hr averages at STP)
  • Measurement range : CO₂: 380–2000 ppm (guaranteed); CH₄: 1.8–500 ppm (guaranteed); H₂O: 0–2.4% (guaranteed)
  • Measurement rate : CO₂-only: ~3 sec; CH₄-only: ~3 sec; Combined mode: ~5 sec
  • Probed volume: Gas sample chamber; typical flow rate: ~25 sccm at 760 Torr
  • Throughput : Not applicable for discrete sampling
  • Laser : Tunable diode laser for cavity ring-down spectroscopy
  • Power Requirements and Consumption : 100–240 VAC, 47–63 Hz; <260 W start-up; 125 W steady-state (analyzer), 35 W (pump)
  • Environmental requirements : Operating temperature: +10 to 35°C; Humidity: <99% RH non-condensing
  • Dimensions and Weight: Analyzer: 43 cm W × 46 cm D × 18 cm H; Weight: 25.4 kg (including external pump)