Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Accelerator Mass Spectroscopy
  • Manufacturer: Ionplus
  • Country: Switzerland
  • Available: Active
All specifications

About

Short description

Next-generation low-energy AMS system for multi-isotope analysis. Combines MICADAS technology with the ETH “TANDY” spectrometer design. Features compact design and high automation.

Specifications

Next-generation low-energy AMS system for multi-isotope analysis. Combines MICADAS technology with the ETH “TANDY” spectrometer design. Features compact design and high automation.

  • Status: Active
  • Country: Switzerland
  • Isotopic: Yes
  • Type of measurable elements: Radiocarbon and cosmogenic nuclides
  • Measurable isotopes/elements: Radiocarbon, actinides, and cosmogenic nuclides - ¹⁰Be, ¹⁴C, ²⁶Al, ⁴¹Ca, ¹²⁹I, U, Pu, and other actinides
  • Analyzer type : Accelerator MS
  • Ion source type: Hybrid cesium negative sputter ion source for solids and gases
  • Inlet types: Random access sample changer with 40 positions
  • High Resolution : Yes
  • Multicollector : Yes
  • Collectors qty : 7 Faraday cups on the high-energy side
  • Sensitivity : High sensitivity with helium stripping
  • Abundance sensitivity : <5 × 10⁻¹³ for U
  • Detection limits : Sub-parts per trillion for selected isotopes
  • Precision : <2 ‰ for ¹⁴C/¹²C and ¹³C/¹²C ratios
  • Detector type : High-resolution ΔE-Eres gas ionization chamber
  • Primary beam : Negative cesium sputter
  • Accelerator voltage : 300 kV
  • Geometry type : Achromatic combination of magnetic and electrostatic deflectors
  • Focusing type : Magnetic and electrostatic focusing
  • Magnetic sector radius/ angle : 90° and 110° magnetic deflectors on high-energy side
  • Electric sector radius/ angle : 120° electrostatic deflector
  • Cooling system : Air-cooled system, no cooling water needed
  • Vacuum system : Vacuum insulated high-voltage platform, no SF₆ needed
  • Gas supply: Helium stripping gas
  • Power Requirements and Consumption : Max. 60 kW (Isotope-dependent)
  • Dimensions and Weight: 3.5 m × 7 m × 2 m, 15,000 kg