Power Requirements and Consumption: MS: AC 230 V 15 A (50/60 Hz) single-phase; Rotary pump: AC 200-240 V 12 A; N₂ gas: max 24.4 L/min ≥97% purity; Ar ≥99.99%; Dry air: max 25 L/min; Total N₂ + air: 48 L/min max
Environmental requirements: 18-28 °C; 20-70% RH (no condensation)
Dimensions and Weight: 118.0 × 54.0 × 61.0 cm; 140.0 kg
Analyzer type: Tandem Quadrupole MS
Scan modes: Q1 Scan/SIM, Q3 Scan/SIM, MRM, Precursor ion scan, Product ion scan, Neutral loss scan, Synchronized Survey Scan
Mass range: m/z 2 to 2,000
High resolution: No
Mass resolution/ Mass resolving power: 0.7 u FWHM (adjustable to 0.5 u)
Dynamic range: 7 orders of magnitude
Sensitivity: ESI pos: 1 pg reserpine S/N > 1,500,000:1 (RMS); ESI neg: 1 pg chloramphenicol S/N > 1,500,000:1 (RMS)
Mass accuracy/Stability: 0.1 u / 0.05 u per 24 hr
Scanning speed: Max 30,000 u/sec (0.1 u step: 300,000 data points/sec)
Dwell time: 0.8 ms
MRM transition rate: Up to 555 channels/sec
Polarity switching: 5 ms
Fragmentation: CID via UFsweeper III collision cell