Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Triple Quad LC-MS
  • Manufacturer: Shimadzu
  • Country: Japan
  • Available: Active
All specifications

About

Short description

Triple quadrupole LC-MS system; ultra-fast MRM acquisition; enhanced sensitivity

Specifications

Triple quadrupole LC-MS system; ultra-fast MRM acquisition; enhanced sensitivity

  • Status: Active
  • Country: Japan
  • Vacuum system : Rotary pump (65 m³/hr) + Triple-inlet turbo molecular pump (40, 260, 210 L/sec)
  • Power Requirements and Consumption : MS: AC 230 V 15 A (50/60 Hz) single-phase; Rotary pump: AC 200-240 V 12 A; N₂ gas: max 24.4 L/min ≥97% purity; Ar ≥99.99%; Dry air: max 20 L/min; Total N₂ + air: 25 L/min max
  • Environmental requirements : 18-28 °C; 20-70% RH (no condensation)
  • Dimensions and Weight: 118.0 × 54.0 × 61.0 cm; 140.0 kg
  • Analyzer type : Tandem Quadrupole MS
  • Scan modes : Q1 Scan/SIM, Q3 Scan/SIM, MRM, Precursor ion scan, Product ion scan, Neutral loss scan, Synchronized Survey Scan
  • Mass range : m/z 2 to 2,000
  • High resolution : No
  • Mass resolution/ Mass resolving power : 0.7 u FWHM (adjustable to 0.5 u)
  • Dynamic range : 7 orders of magnitude
  • Sensitivity : ESI pos: 1 pg reserpine S/N > 2,000,000:1 (RMS); ESI neg: 1 pg chloramphenicol S/N > 2,000,000:1 (RMS)
  • Mass accuracy/Stability : 0.1 u / 0.05 u per 24 hr
  • Scanning speed : Max 30,000 u/sec (0.1 u step: 300,000 data points/sec)
  • Dwell time : 0.8 ms
  • MRM transition rate : Up to 555 channels/sec
  • Polarity switching : 5 ms
  • Fragmentation : CID
  • Ion source types (LC-MS) : ESI, APCI (DUIS-ESI/APCI dual source)

Articles

Other company product

LCMS-9030

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: TOF/Q-TOF LC-MS
  • Manufacturer: Shimadzu
  • Country: Japan
  • Available: Active
Detailed