Specifications

  • Category: Other Types of IOMS
  • Technique: Multicollector Mass Spectrometry for Uranium Hexafluoride
  • Manufacturer: Uralpribor
  • Country: Russia
  • Available: Active
All specifications

About

Short description

System for automated control and isotopic analysis of uranium hexafluoride (UF6) Designed for continuous monitoring in industrial nuclear applications. Features high sensitivity and precision with low UF6 consumption.

Specifications

System for automated control and isotopic analysis of uranium hexafluoride (UF6) Designed for continuous monitoring in industrial nuclear applications. Features high sensitivity and precision with low UF6 consumption.

  • Status: Active
  • Country: Russia
  • Isotopic: Yes
  • Type of measurable elements: Uranium (only)
  • Measurable isotopes/elements: Uranium isotopes
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Electron impact
  • Inlet types: Fully automated sample introduction with pressure control
  • Mass range : 1-350 amu
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : R≥750R
  • Multicollector : Yes
  • Collectors qty : 3
  • Sensitivity : Threshold for uranium: 1×10−5 A at 333 amu
  • Abundance sensitivity : Threshold for uranium: 1×10−5 A at shift 2 from 333 amu
  • Precision : ±0.02%
  • Signal stability, relative standard deviation : ±0.02% over 20 minutes
  • Detector type : Faraday detectors and SEM (otional)
  • Acceleration voltage : 8kV
  • Background : Memory effect: ≤1.005
  • Magnetic sector radius/ angle : 20 cm radius
  • Cooling system : Water cooling
  • Vacuum system : Turbo molecular and ion-getter pumps
  • Gas supply: Uranium hexafluoride (UF6)
  • Power Requirements and Consumption : 220 V, 50 Hz, 3-phase; max power: 3 kVA
  • Environmental requirements : Temperature: 10–35°C; Humidity: ≤75%; Pressure: 84–106.7 kPa
  • Dimensions and Weight: Analytical Unit: 2100 × 1800 × 1700 mm; Weight: 1700 kg