Specifications

  • Category: Other Types of IOMS
  • Technique: Multicollector Mass Spectrometry for Uranium Hexafluoride
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Discontinued
All specifications

About

Short description

Retired multicollector mass spectrometer designed for the isotopic analysis of uranium in uranium hexafluoride (UF6).

Specifications

Retired multicollector mass spectrometer designed for the isotopic analysis of uranium in uranium hexafluoride (UF6).

  • Status: Discontinued
  • Country: Germany
  • Isotopic: Yes
  • Type of measurable elements: Uranium (only)
  • Measurable isotopes/elements: Uranium isotopes
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Crossbeam
  • Inlet types: Automated sample and standard introduction system with cryogenic traps
  • Mass range : 1-350 amu
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : ≥750
  • Multicollector : Yes
  • Collectors qty : Four Faraday cups and one Secondary Electron Multiplier (SEM)
  • Sensitivity : UF6 consumption <1 mg/h at 6×10−10 A ion current
  • Abundance sensitivity : <10−5
  • Precision : < 0.02% external precision for 235U / 238U - ratio; < 0.01% (1 σ) for the ratio of 235U / 238 natural Uranium
  • Signal stability, relative standard deviation : <0.01%for isotopic ratio reproducibility
  • Detector type : Faraday detectors and SEM
  • Possible amplifiers : 10¹⁰, 10¹¹, and 10¹² Ω resistors with dynamic range up to 50 V
  • Background : Memory effect factor <1.003
  • Geometry type : Double-focusing stigmatic geometry
  • Focusing type : Magnetic and electrostatic focusing
  • Magnetic sector radius/ angle : 81 cm radius
  • Cooling system : Water-cooled vacuum system and cryogenic traps
  • Vacuum system : High-capacity turbomolecular pumps and ion-getter pumps; cryogenic cooling for sample handling
  • Gas supply: Uranium hexafluoride (UF6)
  • Power Requirements and Consumption : 230/400 V, 50/60 Hz, approx. 11 kVA
  • Environmental requirements : Temperature: 18–30°C; Humidity: 20–70%, non-condensing

Articles

Other company product

Delta V

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Isotope Ratio Mass Spectrometry for Light Stable Isotopes
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Discontinued
Detailed