Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Tribrid LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
All specifications

About

Short description

Tribrid MS with Orbitrap, quadrupole, linear ion trap; high sensitivity; multiple fragmentation modes

Specifications

Tribrid MS with Orbitrap, quadrupole, linear ion trap; high sensitivity; multiple fragmentation modes

  • Status: Active
  • Country: USA
  • Vacuum system : Split-flow turbo pump
  • Power Requirements and Consumption : 200-240 Vac, 15 A; ~3,500 W
  • Environmental requirements : 15-26 °C; 50-80% RH
  • Dimensions and Weight: 127.0 × 76.7 × 70.3 cm, 180 kg
  • Analyzer type : Orbitrap/Quadrupole/Linear Ion Trap MS
  • Scan modes : Full MS, MS/MS, MSn, SIM, tMSn, targeted MS2
  • Mass range : m/z 50 to 6,000 (MS), m/z 50 to 2,000 (MSn)
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : 500,000 FWHM at m/z 200
  • Dynamic range : 4 orders of magnitude
  • Sensitivity : 1 pg reserpine S/N 100:1
  • Mass accuracy/Stability : <1-3 ppm depending on calibration type
  • Scanning speed : Orbitrap MSn up to 20 Hz; ion trap MSn up to 20 Hz
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : 1 sec or less
  • Fragmentation : CID, HCD, ETD
  • Ion source types (LC-MS) : EASY-Spray, Nanospray Flex, HESI-II, APCI

Articles

Other company product

Trace GC Ultra

  • Category: Chromatography
  • Technique: Gas Chromatography
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Discontinued
Detailed