Specifications

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: Triple Quad ICP-MS
  • Manufacturer: Agilent
  • Country: USA
  • Available: Active
All specifications

About

Short description

Triple quadrupole ICP-MS with MS/MS mode for effective interference removal.

Specifications

Triple quadrupole ICP-MS with MS/MS mode for effective interference removal.

  • Status: Active
  • Country: USA
  • Isotopic: Yes
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Analyzer type : Tandem Quadrupole MS
  • Ion source type: Inductively Coupled Plasma
  • Mass range : Q1: 2–260 amu, Q2: 2–275 amu
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : <0.3 amu (MS/MS mode)
  • Sensitivity : Varies by element, e.g., 150 Mcps/ppm for 7Li, 500 Mcps/ppm for 89Y
  • Dynamic range : 10 orders for Standard; 11 orders for Advanced Applications
  • Detection limits : <0.05 ppt for 115In; <0.02 ppt for 238U in no gas mode
  • Precision : <0.2% RSD for isotope ratio precision
  • Signal stability, relative standard deviation : <3% RSD over 20 minutes, <3% RSD over 2 hours (Factory Specifications)
  • Detector type : Dual-mode discrete dynode electron multiplier
  • Oxides (%) : ≤1.5%
  • Reaction cell system : Yes
  • Reaction cell system details: ORS4 system with He, H2, NH3, O2, and other gases for advanced interference control
  • High Matrix Introduction system : Yes
  • Spray chamber : Scott-type double-pass, quartz, temperature range -5 °C to +20 °C
  • Nebulizer : MicroMist, borosilicate glass
  • Peristaltic pump : 10-roller, 3 channels
  • Torch : One-piece quartz, 2.5 mm id injector with ShieldTorch system
  • Injector : 2.5 mm id
  • Interface : Sampling cone (Ni tip standard, optional Pt tip) and skimmer cone (Ni, optional Pt)
  • RF generator frequency : 27 MHz
  • RF generator power : 500 W to 1600 W
  • CRC gas flow : He, H2, NH3, O2, and other reactive/collision gases
  • Dwell time : 3 ms (standard), 0.1 ms (optional) in TRA mode, no settling time required
  • Focusing type : Off-axis Omega lens
  • Cooling system : Water chiller or water recirculator required
  • Vacuum system : Four-stage differential vacuum system with split-flow turbo molecular pumps
  • Gas supply: Argon (99.99% purity), He (99.999%), NH3, O2 (various purities as required)
  • Power Requirements and Consumption : Single Phase, 200-240 V, 50/60 Hz, 30 A
  • Environmental requirements : Temperature: 15–30 °C; Humidity: 20-80% non-condensing
  • Dimensions and Weight: Mainframe: 1060 mm width x 600 mm depth x 595 mm height; Weight: 139 kg

Articles

Other company product

Agilent 5900

  • Category: Optical Emission Spectroscopy
  • Technique: ICP-OES/ ICP-AES
  • Manufacturer: Agilent
  • Country: USA
  • Available: Active
Detailed