Specifications

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: Triple Quad ICP-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Active
All specifications

About

Short description

High-performance triple quadrupole ICP-MS with advanced interference removal and high sensitivity.

Specifications

High-performance triple quadrupole ICP-MS with advanced interference removal and high sensitivity.

  • Status: Active
  • Country: Germany
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Analyzer type : Tandem Quadrupole MS
  • Ion source type: Inductively Coupled Plasma
  • Mass range : Q1: 2-240 amu, Q3: 2-290 amu
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : User definable in the mass range 2-225 and above m/z 225 only enabled with intelligent Mass Resolution (iMR)
  • Dynamic range : > 11 orders of magnitude (<1 to > 5.10¹⁰ cps)
  • Detection limits : Sub-ppt level; detection limits vary across modes (e.g., <1 ng/L for many elements in He KED mode)
  • Detector type : Dual mode discrete dynode electron multiplier
  • Reaction cell system : Yes
  • Reaction cell system details: QCell with multiple gas modes: He KED, TQ-O2, and TQ-NH3 for enhanced interference removal
  • High Matrix Introduction system : Yes
  • Spray chamber : Glass cyclonic, Peltier-cooled to 2.7 °C
  • Nebulizer : PFA concentric (400 µL·min-1)
  • Peristaltic pump : 12-roller, fourchannel minipump with inert rollers
  • Torch : Quartz torch, demountable
  • Injector : 2.5 mm i.d., quartz injector
  • Interface : Nickel sampler and skimmer cones
  • RF generator frequency : ~27 MHz
  • CRC gas flow : max. 10 mL·min-2
  • Dwell time : of 100 μs in both analog and counting detection modes
  • Cooling system : Water chiller or water recirculator required, Integrated Peltier cooling for the spray chamber
  • Vacuum system : Three stage, differential vacuum with Single split flow turbo molecular pump
  • Gas supply: Argon (99.99% purity), Helium, Oxygen, and Ammonia
  • Power Requirements and Consumption : 200–240 V AC, 50/60 Hz single phase; Apparent: 2,200 VA; Effective: 2,050 W
  • Environmental requirements : 15–35 °C; Rate of change <2 °C·h-1; Humidity Range 20–80% (non-condensing)
  • Dimensions and Weight: 868 mm × 743 mm × 1102 mm, 138 kg

Articles

Other company product

Uranus

  • Category: Other Types of IOMS
  • Technique: Multicollector Mass Spectrometry for Uranium Hexafluoride
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Discontinued
Detailed