Specifications

  • Category: Chromatography
  • Technique: Gas Chromatography
  • Manufacturer: Shimadzu
  • Country: Japan
  • Available: Active
All specifications

About

Short description

Energy-saving GC designed for capillary analysis. Compact and eco-friendly with reduced power and gas consumption.

Specifications

Energy-saving GC designed for capillary analysis. Compact and eco-friendly with reduced power and gas consumption.

  • Status: Active
  • Country: Japan
  • Column type : Capillary
  • Column type : Packed
  • Available detectors : Flame Ionization (FID)
  • Available detectors : Thermal Conductivity (TCD)
  • Available detectors : Electron Capture (ECD)
  • Available detectors : Nitrogen-Phosphorus (NPD)
  • Available detectors : Flame Photometric (FPD)
  • FID details : Max Temp: 400°C, MDL: 2 pg C/s, Linearity: 107
  • TCD details : Max Temp: 350°C, Sensitivity: 9 µV/ppm
  • ECD details : Max Temp: 400°C, MDL: <0.05 pg perchloroethylene
  • NPD details : Max Temp: 450°C, MDL: 5 • 10⁻¹³ g N/s, 5 • 10⁻¹⁴ g P/s
  • FPD details : Max Temp: 400°C, Sensitivity: 2 pg C/s
  • Inlet types : SPL, PTV, Direct Injection
  • Injector general parameters: Max Temp: 400°C, Pressure: 0-970 kPa
  • Flow control : Advanced Flow Controller (AFC)
  • Autosampler compatibility : AOC-20 Series
  • Data acquisition and control: GCsolution Ver.2
  • Column oven specifications : Ambient +10°C to 400°C, Cooling to -50°C with CO₂
  • Safety features : Auto stop/start function, flameout detection
  • Power Requirements and Consumption : 115-230 V, 1700-2600 VA
  • Environmental requirements : Ambient: 10-35°C, Humidity: 20-80%
  • Dimensions and Weight: H: 42 cm, W: 40 cm, D: 49 cm, 30 kg