Specifications

  • Category: Chromatography
  • Technique: Gas Chromatography
  • Manufacturer: Hewlett-Packard
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired GC for routine analyses. Supports various inlets and detectors.

Specifications

Retired GC for routine analyses. Supports various inlets and detectors.

  • Status: Discontinued
  • Country: USA
  • Column type : Capillary
  • Column type : Packed
  • Available detectors : Flame Ionization (FID)
  • Available detectors : Thermal Conductivity (TCD)
  • Available detectors : Electron Capture (ECD)
  • Available detectors : Nitrogen-Phosphorus (NPD)
  • Available detectors : Flame Photometric (FPD)
  • FID details : Max Temp: 450°C, MDL: <5 pg C/s, Linear range: 10^7
  • TCD details : Max Temp: 400°C, MDL: <400 pg/mL, Linear range: 10^5
  • ECD details : Max Temp: 400°C, MDL: <0.04 pg/s lindane, Linear range: 10^4
  • NPD details : Max Temp: 400°C, MDL: <0.4 pg N/s, <0.2 pg P/s
  • FPD details : Max Temp: 300°C, MDL: 20 pg S/s, 0.9 pg P/s
  • Inlet types : Split/Splitless , Packed, On-Column, Split/Splitless
  • Injector general parameters: Max Temp: 400°C, Pressure: 0-100 psi
  • Flow control : Manual
  • Autosampler compatibility : Yes (manual injection)
  • Data acquisition and control: RS-232, INET Communications, Analog
  • Column oven specifications : 4°C above ambient to 450°C, 3 ramps
  • Safety features : Over-temperature protection
  • Power Requirements and Consumption : 120-240 V, 2000 VA
  • Environmental requirements : Ambient: 5-45°C, Humidity: 5-90%
  • Dimensions and Weight: H: 50 cm, W: 60 cm, D: 55 cm, 60 kg

Articles

Other company product

HP 4500

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: Single Quad ICP-MS
  • Manufacturer: Hewlett-Packard
  • Country: USA
  • Available: Discontinued