Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: TOF/Q-TOF LC-MS
  • Manufacturer: Agilent
  • Country: USA
  • Available: Active
All specifications

About

Short description

TOF LC-MS with automated reference mass correction; supports multiple ion sources for molecular profiling and proteomics.

Specifications

TOF LC-MS with automated reference mass correction; supports multiple ion sources for molecular profiling and proteomics.

  • Status: Active
  • Country: USA
  • Vacuum system : -
  • Power Requirements and Consumption : 200-240 VAC, 2500 VA max, 50/60 Hz, 15 A circuit
  • Environmental requirements : Temperature 15 to 35 °C, maintain 2 ppm accuracy within 2 °C drift/hour
  • Dimensions and Weight: 83.0 × 73.0 × 133.0 cm; 117.9 kg
  • Analyzer type : Time-of-Flight MS
  • Scan modes : -
  • Mass range : m/z 25 to 20,000
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : 22,000 FWHM at m/z 1,522
  • Dynamic range : 5 orders of magnitude
  • Sensitivity : MS: >50:1 with AJS, >10:1 with ESI Option 200 at 1 pg
  • Mass accuracy/Stability : <1 ppm accuracy; 2 ppm within 2 °C drift per hour
  • Scanning speed : MS: 30 spectra/sec
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : < 1 s
  • Fragmentation : -
  • Ion source types (LC-MS) : ESI, APCI, APPI, Multimode (ESI+APCI), HPLC-Chip/MS, AP/MALDI PDF