Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: TOF/Q-TOF LC-MS
  • Manufacturer: SCIEX
  • Country: USA
  • Available: Active
All specifications

About

Short description

Zeno trap-enabled QTOF. ZT Scan DIA and SWATH supported. Electron-activated dissociation available.

Specifications

Zeno trap-enabled QTOF. ZT Scan DIA and SWATH supported. Electron-activated dissociation available.

  • Status: Active
  • Country: USA
  • Vacuum system : -
  • Power Requirements and Consumption : -
  • Environmental requirements : -
  • Dimensions and Weight: -
  • Analyzer type : Quadrupole/Time-of-Flight MS
  • Scan modes : SWATH DIA; ZT Scan DIA (7600+); DDA/IDA; targeted (MRMHR).
  • Mass range : 40 to 40 kDa in TOF
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : ≥ 42,000 at m/z 956.
  • Dynamic range : 5 orders of magnitude
  • Sensitivity : 5-20× MS/MS sensitivity gain with Zeno trap; >90% ions injected into TOF (duty cycle).
  • Mass accuracy/Stability : <2 ppm RMS external; <1 ppm RMS internal.
  • Scanning speed : MS/MS up to 133 Hz; up to 640 Hz with ZT Scan DIA (750 Da/s, 5 Da sliding window) on 7600+.
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : -
  • Fragmentation : EAD; CID; EAD-CID hybrid noted.
  • Ion source types (LC-MS) : Turbo V (ESI ; APC; Optiflow Turbo V; nano).