Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: TOF/Q-TOF LC-MS
  • Manufacturer: Waters
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired Q-TOF LC-MS for accurate-mass MS and MS/MS; quadrupole precursor selection; suited to broad screening and confirmation.

Specifications

Retired Q-TOF LC-MS for accurate-mass MS and MS/MS; quadrupole precursor selection; suited to broad screening and confirmation.

  • Status: Discontinued
  • Country: USA
  • Vacuum system : Differentially pumped, automated system with air-cooled turbomolecular pumps + rotary or oil-free backing pump; isolation valve; tool-free service access
  • Power Requirements and Consumption : -
  • Environmental requirements : -
  • Dimensions and Weight: 68.7 × 152.0 × 87.3 cm
  • Analyzer type : Quadrupole/Time-of-Flight MS
  • Scan modes : MS scanning; MS/MS product ion scanning; UPLC/FastDDA; UPLC/MSE; Ionization mode switching (ESCi); External contact start/stop/events; Analogue
  • Mass range : Up to m/z 100,000
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : >22,500 FWHM
  • Dynamic range : 4 orders of magnitude
  • Sensitivity : MS sensitivity (ESI+): >6,400 cps (m/z 556, 50 pg/µL leucine enkephalin, 22,500 FWHM); MS sensitivity (ESI−): >7,200 cps (m/z 503, 500 pg/µL raffinose, 22,500 FWHM)
  • Mass accuracy/Stability : <3 ppm RMS with lock mass (10 consecutive raffinose measurements at m/z 527.1588)
  • Scanning speed : -
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : -
  • Fragmentation : CID (T-Wave collision cell)
  • Ion source types (LC-MS) : ZSpray™ dual-orthogonal API; Multi-mode source ESI/APCI/ESCi (optional); APCI probe (optional); APPI/APCI dual mode (optional)