Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: TOF/Q-TOF LC-MS
  • Manufacturer: JEOL
  • Country: Japan
  • Available: Active
All specifications

About

Short description

Retired high-throughput mass spectrometer with multiple ionization options, including DART and ColdSpray; maintenance-accessible design.

Specifications

Retired high-throughput mass spectrometer with multiple ionization options, including DART and ColdSpray; maintenance-accessible design.

  • Status: Active
  • Country: Japan
  • Vacuum system : -
  • Power Requirements and Consumption : Main console: Single phase AC 190-210 V or 220-240 V, 50-60 Hz, 20 A
  • Environmental requirements : Room temperature: 20-27 °C; Humidity: 30-70% (no condensation); Floor vibration: Amplitude (p-p) 25 μm or less, acceleration 0.1 m/s² or less
  • Dimensions and Weight: MS: 69.0 × 90.5 × 115.6 cm, 305 kg
  • Analyzer type : Time-of-Flight MS
  • Scan modes : -
  • Mass range : Up to m/z 10,000
  • High resolution : No
  • Mass resolution/ Mass resolving power : ≥10,000 at m/z 609
  • Dynamic range : -
  • Sensitivity : -
  • Mass accuracy/Stability : ≤3 ppm
  • Scanning speed : Up to 50 spectra/sec
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : -
  • Fragmentation : -
  • Ion source types (LC-MS) : DART, ESI, APCI, ColdSpray

Articles

Other company product

CRYO ARM 200 II (JEM-Z200FSC)

  • Category: Electron Microscopy
  • Technique: Transmission Electron Microscopy
  • Manufacturer: JEOL
  • Country: Japan
  • Available: Active
Detailed