Specifications

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: ICP-TOF-MS
  • Manufacturer: GBC Scientific Equipment
  • Country: Australia
  • Available: Active
All specifications

About

Short description

The third generation bench top ICP-TOFMS from GBC with Octopole Collision Cell (OCC) technology for interference management.

Specifications

The third generation bench top ICP-TOFMS from GBC with Octopole Collision Cell (OCC) technology for interference management.

  • Status: Active
  • Country: Australia
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Analyzer type : Time-of-Flight MS
  • Ion source type: Inductively Coupled Plasma
  • Mass range : 1-260 amu
  • Mass resolution/ Mass resolving power : 0,4 amu
  • Detection limits : ppt
  • Detector type : Discrete dynode electron multiplier (DDEM)
  • Reaction cell system : Yes
  • Reaction cell system details: Octopole Collision Cell (OCC)
  • Spray chamber : thermostatted glass jacketed spray chamber
  • Peristaltic pump : four-channel 12-roller pump, speed 0-60 rpm
  • Torch : Low flow, low power, single piece quartz torch
  • Interface : Threecone watercooled interface
  • RF generator frequency : 27.12 MHz
  • RF generator power : 500W - 1500W
  • Vacuum system : Differential vacuum system with turbomolecularpumps and rotary vane backing pump
  • Gas supply: Argon and additional reactive gases

Articles

Other company product

Integra 6000

  • Category: Optical Emission Spectroscopy
  • Technique: ICP-OES/ ICP-AES
  • Manufacturer: GBC Scientific Equipment
  • Country: Australia
  • Available: Active
Detailed