Specifications

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: Single Quad ICP-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Active
All specifications

About

Short description

Single quadrupole ICP-MS designed for high sensitivity and matrix robustness.

Specifications

Single quadrupole ICP-MS designed for high sensitivity and matrix robustness.

  • Status: Active
  • Country: Germany
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Analyzer type : Quadrupole MS (single)
  • Ion source type: Inductively Coupled Plasma
  • Mass range : 2-290 amu
  • Mass resolution/ Mass resolving power : User definable in the mass range 2-225 and above m/z 225 only enabled with intelligent Mass Resolution (iMR)
  • Sensitivity : kcps/ppb: 7Li 80; 59Co 290; 115In 600; 209Bi 500; 238U 850
  • Abundance sensitivity : < 0.5 ppm at m-1 and m+1 (m = 133Cs)
  • Dynamic range : >11 orders of magnitude (<1 to >5·10e10 cps)
  • Detection limits : Sub-ppt level, varies by element; e.g., <0.5 ng/L for many elements in He KED mode
  • Precision : 107Ag/109Ag <0.1% RSD
  • Signal stability, relative standard deviation : RSD %: Short termc <2 (10 min); Long term <3 (2 h)
  • Detector type : Dual mode discrete dynode electron multiplier
  • Oxides (%) : 140Ce16O/140Ce <2.0%; 140Ce++/140Ce+ <3.0%
  • Reaction cell system : Yes
  • Reaction cell system details: QCell with Kinetic Energy Discrimination (KED) and low-mass cutoff using He gas
  • High Matrix Introduction system : Yes
  • Spray chamber : Glass spray chamber, cooled at 2.7 °C
  • Nebulizer : PFA concentric (400 µL·min-1)
  • Peristaltic pump : 12-roller, fourchannel minipump with inert rollers
  • Torch : Quartz torch, demountable
  • Injector : 2.5 mm i.d., quartz injector
  • Interface : Nickel-based sampler and skimmer cones
  • RF generator frequency : ~27 MHz
  • CRC gas flow : max. 10 mL·min-1
  • Dwell time : of 100 μs in both analog and counting detection modes
  • Cooling system : Water chiller or water recirculator required, Integrated Peltier cooling for the spray chamber
  • Vacuum system : Three stage, differential vacuum with Single split flow turbo molecular pump
  • Gas supply: Argon (99.99% purity), Helium (99.999% purity)
  • Power Requirements and Consumption : 200–240 V AC, 50/60 Hz single phase; Apparent: 2,200 VA; Effective: 2,050 W
  • Environmental requirements : 15–35 °C; Rate of change <2 °C·h-1; Humidity Range 20–80% (non-condensing)
  • Dimensions and Weight: 868 mm × 743 mm × 1102 mm, 131 kg

Articles

Other company product

Talos F200C

  • Category: Electron Microscopy
  • Technique: Transmission Electron Microscopy
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
Detailed