Specifications

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: Single Quad ICP-MS
  • Manufacturer: Shimadzu
  • Country: Japan
  • Available: Active
All specifications

About

Short description

Compact ICP-MS with low running costs.

Specifications

Compact ICP-MS with low running costs.

  • Status: Active
  • Country: Japan
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Analyzer type : Quadrupole MS (single)
  • Ion source type: Inductively Coupled Plasma
  • Mass range : 5 - 260 amu
  • Mass resolution/ Mass resolving power : 0.6–0.8 amu
  • Dynamic range : 9 orders of magnitude
  • Detection limits : ppt-level sensitivity
  • Signal stability, relative standard deviation : <2% for short-term, <3% for long-term stability
  • Detector type : Secondary Electron Multiplier Tube Detector
  • Reaction cell system : Yes
  • Reaction cell system details: Collision cell with helium for polyatomic interference reduction
  • Spray chamber : Cyclone chamber, cooled by a Peltier element
  • Nebulizer : Coaxial nebulizer with high aerosol efficiency
  • Torch : Mini-torch design for reduced argon consumption
  • Interface : Tool-free, easily removable interface for rapid maintenance
  • Focusing type : Lens system located behind the collision cell for improved ion transmission
  • Cooling system : Requires a cooling water circulator
  • Vacuum system : Compact, three-stage split-flow turbomolecular pump
  • Gas supply: Argon, Helium
  • Power Requirements and Consumption : Single-phase 200–240 V ± 10 %, 50/60 Hz, 6 kVA
  • Environmental requirements : Temperature: 18–28 °C; Humidity: 20–70% RH
  • Dimensions and Weight: 870 mm (width) x 645 mm (depth) x 587 mm (height); Weight: 140 kg