Specifications

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: Triple Quad ICP-MS
  • Manufacturer: Agilent
  • Country: USA
  • Available: Active
All specifications

About

Short description

ICP-MS system for high-throughput analysis, handling up to 25% solids in samples.

Specifications

ICP-MS system for high-throughput analysis, handling up to 25% solids in samples.

  • Status: Active
  • Country: USA
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Analyzer type : Quadrupole MS (single)
  • Ion source type: Inductively Coupled Plasma
  • Mass range : 2–260 amu
  • Mass resolution/ Mass resolving power : Variable from 0.3 u to 1.0 amu
  • Sensitivity : Varies by element, e.g., 110 Mcps/ppm for 7Li, 540 Mcps/ppm for 238U
  • Dynamic range : 10 orders of magnitude (0.3 cps to 4 Gcps)
  • Detection limits : <0.1 ppt for 115In, <0.04 ppt for 209Bi
  • Precision : <0.1% RSD for isotope ratio precision
  • Signal stability, relative standard deviation : <3% RSD over 20 minutes, <3% RSD over 2 hours (Factory Specifications)
  • Detector type : Dual-mode discrete dynode electron multiplier
  • Oxides (%) : ≤1.5%
  • Reaction cell system : Yes
  • Reaction cell system details: Octopole Reaction System (ORS) with helium collision mode; optional H2 gas line
  • High Matrix Introduction system : Yes
  • Spray chamber : Scott-type double-pass, quartz, temperature range -5 °C to +20 °C
  • Nebulizer : MicroMist, borosilicate glass
  • Peristaltic pump : 10-roller, 3 channels
  • Torch : One-piece quartz, 2.5 mm id injector with ShieldTorch system
  • Injector : 2.5 mm id
  • Interface : Sampling cone (Ni-tipped standard, optional Pt-tipped) and skimmer cone (Ni, optional Pt-tipped)
  • RF generator frequency : 27 MHz
  • RF generator power : 500 W to 1600 W
  • CRC gas flow : Helium and optional hydrogen
  • Dwell time : 3 ms (standard), 0.1 ms (optional) in TRA mode, no settling time required
  • Focusing type : Off-axis Omega lens
  • Cooling system : Water chiller or water recirculator required
  • Vacuum system : Three-stage differential vacuum system with turbo molecular and rotary pumps
  • Gas supply: Argon (99.99% purity), Helium (99.999% purity), H2 (99.999% purity)
  • Power Requirements and Consumption : Single Phase, 200-240 V, 50/60 Hz, 30 A
  • Environmental requirements : Temperature: 15–30 °C; Humidity: 20-80% non-condensing
  • Dimensions and Weight: Mainframe: 730 mm width x 600 mm depth x 595 mm height; Weight: 100 kg

Articles

Other company product

Agilent 5100

  • Category: Optical Emission Spectroscopy
  • Technique: ICP-OES/ ICP-AES
  • Manufacturer: Agilent
  • Country: USA
  • Available: Discontinued
Detailed