Retired Single Quad GC-MS with ExtractaBrite Ion Source.
Specifications
Retired Single Quad GC-MS with ExtractaBrite Ion Source.
Status: Discontinued
Country: USA
Available configurations: EI; EI/CI
Analyzer type: Quadrupole MS (single)
Scan modes: Full scan, SIM
Mass range: 1.2 to 1100 amu
Mass resolution/ Mass resolving power: Unit mass
Ionization Modes: Electron (EI)
Ionization Modes: Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
Sensitivity: 1 pg OFN S/N ≥450
Mass accuracy/Stability: ±0.1 amu/48h
Detector type: DynaMax XR Detection System
Electron energy: 0–150 eV
Scanning speed: Up to 11,111 u/sec
Direct inlet option: Yes
Installation specifications: Electron Ionization 1 μL of 1 pg/μL octafluoronaphthalene (OFN) will produce the following minimum signal to noise for m/z 272 when scanning from 50 – 300 u: 450:1 using helium as carrier gas
Positive Chemical Ionization 1 μL of 100 pg/μL benzophenone will produce the following minimum signal to noise for m/z 183 when scanning from 80 – 230 u using methane reagent gas: 300:1 using helium as carrier gas
Negative Chemical Ionization 2 μL of 100 fg/μL of OFN will produce the following minimum signal to noise for m/z 272 when scanning from 50 – 300 u using methane reagent gas: 600:1 using helium as carrier gas