Specifications

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: Multicollector ICP-MS
  • Manufacturer: Nu Instruments
  • Country: UK
  • Available: Active
All specifications

About

Short description

A third-generation multi-collector ICP-MS with Enhanced Abundance Transfer Optics (EATO) for improved sensitivity and a robust RF generator with frequency tuning.

Specifications

A third-generation multi-collector ICP-MS with Enhanced Abundance Transfer Optics (EATO) for improved sensitivity and a robust RF generator with frequency tuning.

  • Status: Active
  • Country: UK
  • Isotopic: Yes
  • Elemental: Yes
  • Type of measurable elements: Metals and non-metals
  • Measurable isotopes/elements: Lithium to uranium, including actinides and heavy metals
  • Analyzer type : Magnetic Sector MS
  • Ion source type: Inductively Coupled Plasma
  • Inlet types: Laser ablation, desolvation systems, autosamplers, gas chromatography, and cold vapor generation
  • Mass range : 3–310 amu
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : >800 (typical), with pseudo-high-resolution and full high-resolution capabilities
  • Multicollector : Yes
  • Collectors qty : 16 Faraday cups (basic) with up to 6 ion counters (maximum)
  • Sensitivity : Dry plasma: U: 410 V/ppm, Pb: 500 V/ppm Wet plasma: U: 40 V/ppm, Pb: 50 V/ppm
  • Abundance sensitivity : <2ppm at -1 amu spacing relative to 238U; <0.5 ppm with deceleration filter
  • Dynamic range : >10¹² with enhanced collector and amplifier configurations
  • Detection limits : Sub-ppq for trace isotopes
  • Precision : High precision for isotopic ratio analysis; <0.032% RSD for Nd isotopes over 7 hours
  • Signal stability, relative standard deviation : <0.05 ppm with EATO and deceleration filters
  • Detector type : Faraday detectors, discrete dynode SEMs, and Daly detectors
  • Possible amplifiers : 10¹⁰, 10¹¹, 10¹², and 10¹³ ohm resistors with software-switchable configurations
  • Acceleration voltage : 10 kV
  • Reaction cell system : No
  • Spray chamber : Peltier-cooled cyclonic spray chamber
  • Nebulizer : Concentric nebulizer
  • Peristaltic pump : multi-channel for sample introduction and drainage
  • Torch : Advanced torch box design for flexibility and stability
  • Interface : High-capacity Jet Interface with optimized pumping for vacuum stability
  • RF generator frequency : 27.12 MHz
  • RF generator power : 2.0 kW
  • Geometry type : Nier-Johnson forward geometry
  • Focusing type : Double focusing with magnetic and electrostatic sectors
  • Magnetic sector radius/ angle : 250 mm/70 deg
  • Electric sector radius/ angle : 350 mm/ 70 deg
  • Cooling system : Water-cooled system for optimal thermal stability
  • Vacuum system : Fully turbo-pumped with additional rotary pumps for plasma and interface regions
  • Gas supply: Argon with additional helium for laser ablation systems