Specifications

  • Category: Gas Chromatography - Mass Spectrometry
  • Technique: Triple Quad GC-MS
  • Manufacturer: Scion Instruments
  • Country: UK
  • Available: Active
All specifications

About

Short description

Triple Quad GC-MS with CBS MRM Library

Specifications

Triple Quad GC-MS with CBS MRM Library

  • Status: Active
  • Country: UK
  • Country: Netherlands
  • Available configurations: EI; EI/CI
  • Analyzer type : Quadrupole MS (single)
  • Scan modes : Full Scan, Precursor Ion Scan, Product Ion Scan, Neutral Loss Scan, SIM, MRM, Full Scan/SIM, Full Scan/MRM
  • Mass range : 1 to 1200 Da
  • Mass resolution/ Mass resolving power : User adjustable 0.7 – 4 Da
  • Ionization Modes : Electron (EI)
  • Ionization Modes : Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
  • Sensitivity : MRM S/N ≥50,000:1 for 100 fg OFN
  • Mass accuracy/Stability : ±0.1 Da over 48h
  • Detector type : EDRTM Electron multiplier with ±5 kV post acceleration
  • Electron energy : 0–150 eV
  • Scanning speed : Up to 30,000 Da/sec
  • Ion source temperature : 100 – 350°C
  • Maximum column Flow Rate : Up to 25 mL/min
  • Dwell time : 0.5 ms
  • MRM transition rate : Up to 1000 MRM/sec
  • Collision cell : 180° curved collision cell
  • Collision energy : 0 – 75 eV
  • Installation specifications : EI Full Scan 1 pg OFN from m/z 50 to 300 for m/z 272 S/N ≥ 1,500:1; EI SIM 25 fg OFN for m/z 272 S/N ≥ 50:1; EI MRM 100 fg OFN for m/z 272>222 S/N ≥ 50,000:1; EI MRM IDL 8 injections 10 fg OFN RSD < 4 fg; PCI Full Scan 10 pg BZP (80→230) for 183 S/N > 50:1; PCI SIM 1 pg BZP for m/z 183>105 S/N > 50:1; PCI MRM 100 fg BZP for m/z 183>105 S/N > 150:1; NCI Full Scan 1 pg OFN (200→300) for m/z 272 S/N > 4,000:1; NCI MRM 10 fg OFN for m/z 272 S/N > 400:1
  • Vacuum system : Dual-stage turbo pump 310/400 L/s
  • Power Requirements and Consumption : 100-240V, 50/60 Hz ±3 Hz, 1200 VA
  • Environmental requirements : 15 – 33°C, 20 – 80% RH
  • Dimensions and Weight: 45x28x57 cm, 40 kg