Specifications

  • Category: Gas Chromatography - Mass Spectrometry
  • Technique: Triple Quad GC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired Triple Quad MS with HyperQuad Analyzer and Matrix Tolerance.

Specifications

Retired Triple Quad MS with HyperQuad Analyzer and Matrix Tolerance.

  • Status: Discontinued
  • Country: USA
  • Available configurations: EI; EI/CI
  • Analyzer type : Tandem Quadrupole MS
  • Scan modes : Full-scan MS in Q1 or Q3; SRM; High Selective Reaction Monitoring (H-SRM); SIM in Q1 or Q3; High selective Selected Ion Monitoring (H-SIM) mode; Product Ion Scan; Precursor Ion Scan; Neutral Loss Scan
  • Mass range : 10 to 3000 amu
  • Mass resolution/ Mass resolving power : 0.4 to 5.0 Da adjustable
  • Ionization Modes : Electron (EI)
  • Ionization Modes : Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
  • Mass accuracy/Stability : ±0.002 amu/24h
  • Detector type : Off-axis Electron Multiplier
  • Electron energy : 0–140 eV
  • Scanning speed : Up to 5,000 Da/sec
  • Direct inlet option : Yes
  • Ion source temperature : 125 to 350°C
  • Dwell time : 1 ms
  • MRM transition rate : Up to 3000 MRM/run
  • Collision cell : 90° collision cell
  • Collision energy : 0-200 eV
  • Installation specifications : Electron Ionization SRM: Hot splitless injection of 1 µL of 100 fg/µL octafluoronaphthalene (OFN) in iso-octane for m/z 271.99>240.99 S/N > 2,500:1, SRM mode at 5 scans/sec, mass resolution 0.7 Da FWHM
  • Vacuum system : Turbo Pump 270 L/s
  • Power Requirements and Consumption : 230V, 30A, 50/60 Hz,
  • Environmental requirements : 15-27°C, 40-80% RH
  • Dimensions and Weight: 69x56x79 cm, 118 kg