Retired Triple Quad MS with HyperQuad Analyzer and Matrix Tolerance.
Specifications
Retired Triple Quad MS with HyperQuad Analyzer and Matrix Tolerance.
Status: Discontinued
Country: USA
Available configurations: EI; EI/CI
Analyzer type: Tandem Quadrupole MS
Scan modes: Full-scan MS in Q1 or Q3; SRM; High Selective Reaction Monitoring (H-SRM); SIM in Q1 or Q3; High selective Selected Ion Monitoring (H-SIM) mode; Product Ion Scan; Precursor Ion Scan; Neutral Loss Scan
Mass range: 10 to 3000 amu
Mass resolution/ Mass resolving power: 0.4 to 5.0 Da adjustable
Ionization Modes: Electron (EI)
Ionization Modes: Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
Mass accuracy/Stability: ±0.002 amu/24h
Detector type: Off-axis Electron Multiplier
Electron energy: 0–140 eV
Scanning speed: Up to 5,000 Da/sec
Direct inlet option: Yes
Ion source temperature: 125 to 350°C
Dwell time: 1 ms
MRM transition rate: Up to 3000 MRM/run
Collision cell: 90° collision cell
Collision energy: 0-200 eV
Installation specifications: Electron Ionization SRM: Hot splitless injection of 1 µL of 100 fg/µL octafluoronaphthalene (OFN) in iso-octane for m/z 271.99>240.99 S/N > 2,500:1, SRM mode at 5 scans/sec, mass resolution 0.7 Da FWHM
Vacuum system: Turbo Pump 270 L/s
Power Requirements and Consumption: 230V, 30A, 50/60 Hz,