Retired Triple Quad MS with Ultra Selectivity and HyperQuad Technology.
Specifications
Retired Triple Quad MS with Ultra Selectivity and HyperQuad Technology.
Status: Discontinued
Country: USA
Available configurations: EI; EI/CI
Analyzer type: Tandem Quadrupole MS
Scan modes: Full-scan MS in Q1 or Q3; SRM; Ultra high Selective Reaction Monitoring (U-SRM); SIM in Q1 or Q3; Ultra high selective Selected Ion Monitoring (U-SIM) mode; Product Ion Scan; Precursor Ion Scan; Neutral Loss Scan
Mass range: 10 to 1500 amu
Mass resolution/ Mass resolving power: 0.1 to 5.0 Da adjustable
Ionization Modes: Electron (EI)
Ionization Modes: Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
Sensitivity: IDL ≤6 fg OFN
Mass accuracy/Stability: ±0.1 amu/24h
Detector type: Off-axis Electron Multiplier
Electron energy: 0–140 eV
Scanning speed: Up to 5,000 Da/sec
Direct inlet option: Yes
Ion source temperature: 125 to 350°C
Dwell time: 1 ms
MRM transition rate: Up to 3000 MRM/run, Fast MRM scan speed of more than 300 SRM/s
Collision cell: 90° collision cell
Collision energy: 0-200 eV
Installation specifications: EI SRM (U-SRM): 100 fg OFN for m/z 271.99>240.99 S/N ≥ 12,000:1, Q1 at 0.2 Da FWHM, Q3 at 0.7 Da FWHM, scan rate ≥ 5 scans/sec;
PCI SRM: 100 fg BZP(d10) for m/z 193.14>110.07 S/N ≥ 300:1, Q1 at 0.7 Da FWHM, Q3 at 0.7 Da FWHM, scan rate ≥ 5 scans/sec, methane as reagent gas
Vacuum system: Turbo Pump 270 L/s
Power Requirements and Consumption: 230V, 30A, 50/60 Hz